![NewAE CHIPSHOUTER CW520 Скачать руководство пользователя страница 58](http://html.mh-extra.com/html/newae/chipshouter-cw520/chipshouter-cw520_user-manual_1678377058.webp)
ChipSHOUTER Users Manual
: Serial Interface
58
arm
a
Arms device (charges high voltage capacitor bank).
If no trigger occurs the device will automatically
disarm after
arm_timeout
seconds.
If arming fails, the device may have an active
fault. Check active and latched faults with the
get fault command.
Arm
A
Clears latched faults and arms device, equivalent
to running
set fault none
followed by
arm
. This
command is useful when using the external trigger,
as you may need to quickly clear a latched fault
and arm the device.
disarm
d
Disarms device (turns off high voltage and dis-
charges capacitor bank internally).
safe
S
See disarm.
pulse
p
Triggers simple or pattern pulse according to set-
tings.
Reset_config
Reset
Resets configuration to product default, will
cause EEPROM CRC error on next boot.
reset
Reboots the board, maintains most settings.