Vidia-Discrete
User Safety • 7
Figure 2:
Certification label
on the back of the
laser
Using the Safety Interlock
The safety interlock connector on the back of the controller is provided
for external safety systems. The system is shipped with a jumper across
the interlock terminals. Do not remove this jumper unless you are
using the safety interlock feature; the laser will not emit light unless the
interlock circuit is closed. The circuit carries 15-V DC.
Interlock
Laser
Sync
Output
Current
Modulation
Input
Detector
Input
IEEE 488
RS232
120Vac
WARNING
For continued protection against fire hazard, replace
only with the same type and rating of fuse.
Input VAC
90 - 120
200 - 250
CAUTION
Risk of electric shock, do not remove
cover. Refer servicing to qualified personnel.
10 - 250 VAC
1.2 - 0.5A, 48-66Hz
Fuse
2.0 A T
1.0 A T
5215 Hellyer Ave.
■
San Jose
■
CA 95051
-0905
U.S. Patent #5,319,668 and Patents Pending
Model Number:
Serial Number:
Manufactured:
This product conforms to the applicable requirements of
21 CFR 1040.10 and 1040.11 at the date of manufacture.
5215 Hellyer Ave.
■
San Jose
■
CA 95138-1001
U.S. Patent #5,319,668 and Patents Pending
Model Number:
Serial Number:
Manufactured:
This product conforms to the applicable requirements of
21 CFR 1040.10 and 1040.11 at the date of manufacture.
64X7 rev E.fm Page 7 Tuesday, April 3, 2001 4:16 PM