- 11 -
8. Troubleshooting
Before requesting for repairs, please walk through the Troubleshooting below
Problem Cause
Remedy
No indication appears on
the display when the
selector switch is set to
“BATT” or “MEAS”.
Battery life.
Battery polarity is
reversed.
Remove the batteries and insert
again (page 4).
Insert batteries correctly with
respect to their polarity (page 4).
The battery alarm lamp is
lit or blinking when
selector switch is set to
“BATT” or “MEAS”.
Battery life.
Replace batteries (page 4).
The reading of the display
does not stabilize in the
measurement.
The temperature of the
grease sample is too
high to measure.
The sampling port
becomes warm.
Lower the temperature of the
grease sample and measure.
Hold a finger grip of sampling
container and measure again.