6. Error Message
This instrument has a function of indicating error messages.
Read along with “
8. Troubleshooting
” on page 11 to take measures.
Error Message
Cause
Measure
Blinking
[ - - - - ]
Zero adjustment could not be performed,
because the power was turned on while
the syringe holder was inserted.
Remove the syringe
holder and confirm the
indication becomes 0
before using.
Measurement error.
The syringe holder was removed before
the measurement is completed.
Remove the syringe
holder again and restart
measurement.
[ P. OFF ]
The power supply was forced to shut
down due to the power supply voltage
drops.
Turn the selector switch
to “OFF” and then turn to
“BATT” or “MEAS” again.
[ Err. E ]
Malfunction.
Writing error of non-volatile memory.
Turn the selector switch
to “OFF” and then turn to
“BATT” or “MEAS” again.
If the indication is still the
same, it is in need of
repair.
[ Err. r ]
Malfunction.
Reading error of non-volatile memory.
[ Err. S ]
Malfunction.
Sum check error of non-volatile memory.
[ Err. A ]
Malfunction.
There is no response from AD converter.