Multitone Audio Test System
User Manual
V 3.32
45 / 71
INPut:SYNC [INTernal¦INTNoheader¦EXTernal¦EXTNoheader]
Use
Defines the synchronization mode of the instrument.
Range
INT ¦ INTN ¦ EXT ¦ EXTN
boolean
INT
INTN
EXT
EXTN
Analyzer sampling clock is derived from the generator crystal. No frequency shift
correction is performed. Use this mode if analyzer and generator are located
together and no frequency shifts are expected. Max. allowable time delay is 1s.
Analyzer sampling clock is derived from the generator crystal. No frequency shift
during transmission must occur. The multitone signal is sent out without any
header. The analyzer expects the multitone signal without trigger and SYNC
block. Use this mode if analyzer and generator are located together but no header
information can be transmitted (e.g. muted measurements) or for analysis of
externally generated signals. Max. allowable time delay is 50ms.
The analyzer clock is synchronized to the frequency of the SYNC block in the
header of the received multitone burst. Frequency shifts are compensated. This
mode is recommended if notable frequency shifts are expected. Max. allowable
time delay is 1s.
No synchronization is transmitted or performed at all. Analyzer clock runs at the
frequency synchronized to the last transmitted multitone burst in Sync Mode
EXT. No further tuning will be performed. Analyzer expects the signal with no
trigger information and generator is set to transmit the multitone signal only. No
major time delay must occur in that mode. This mode requires at least one
measurement in EXT mode before to ensure that the analyzer crystal is tuned.
Default
INT
Query
Use command INPut[1-2]:STATus?
Example
:
→
:
→
Inp:Sync Internal
INPUT:SYNC INTN
Explanation
Any multitone burst transmission & sampling must be initiated by OUTPut:MTONe:STARt.
INPut:SWFilter [OFF¦CWE¦CCITT]
Use
Activates one of the implemented software weighting filters. Filters are selected for both
channels simultaneously. The filters may be engaged also after the measurements has been
performed. This allows to get first unweighted and afterwards weighted results.
Range
OFF ¦ CWE ¦ CCITT
boolean
Default
OFF
Query
Use command INPut[1-2]:STATus?
Example
:
→
:
→
INPUT:SWFILTER OFF
Inp:SWF CWE