Multitone Audio Test System
User Manual
V 3.32
11 / 71
2 I
NSTALLATION
This chapter is intended as help for proper unpacking and installation of the
>@9
system.
Please read it carefully to avoid wrong connections or inconveniences during operation of
the instrument.
Unpacking
>@9
has been carefully packed by NEUTRIK to avoid damages during transportation.
Should the box show severe damages, please immediately check the instrument inside on
external impacts. In case of any visible damage, please do not send the instrument back but
contact your local dealer and / or the carrier to avoid loss of claims for replacement.
Rack Mount
>@9
is designed to mount in a 19“ Rack and occupies one unit of height or rack space
(1.75“) only. Please allow at least 2“ additional depth at the rear side for all necessary
connectors. Make sure there is enough air circulation around the unit for cooling purposes
and please do not place
>@9
besides high temperature devices such as power amplifiers in
order to avoid overheating.
The specified operating temperature ranges between 5° and 45°C (40-110F) while humidity
must not exceed 90% R.H. non-condensing.
AC Power Connection
Before connecting the instrument via mains cable to the power source, make sure that the
voltage selector label on the connector / fuse holder assembly of the
>@9
system matches
the supply voltage of the local area. If the instrument is not compatible with the available
power source, follow the next paragraph to change the voltage selector.
>@9
can operate from 100VAC, 120VAC and
230VAC sources. To reconfigure the input line
voltage, remove the power cable and open the
flap of the connector/fuse holder at the rear
side of
>@9
. Either press a small screwdriver
into the slot to open the flap as shown in
Fig. 1
or ruin your fingernails.
Take out the drum and insert it in the new position so that the matching voltage indication
points towards you. At the same time replace the mains fuse with the proper current rating.
For voltages of 100V to 120V a slow 2A fuse has to be installed, while for 230V a slow 1A
fuse is appropriate.
After selection of the correct mains voltage and fuse, close the flap and insert the power
cable.
>@9
is designed with a protective ground (earth) connection through
the ground wire of the power cord. This connection is essential for
safe operation. Never operate the instrument if safety ground is
unavailable or has been compromised.
Fig. 1 Voltage Selector