a.  Field strengths from fixed transmitters, such as base stations for radio 

(cellular/cordless) telephones and land mobile radios, amateur radio, 
AM and FM radio broadcast and TV broadcast cannot be predicted 
theoretically with accuracy. To assess the electromagnetic environment 
due to fixed RF transmitters, an electromagnetic site survey should 
be considered. If the measured field strength in the location in which 
this device is used exceeds the applicable RF compliance level above, 
this device should be observed to verify normal operation. If abnormal 
performance is observed, additional measures may be necessary, such 
as re-orienting or relocating this device.

b. 

than 3V/m.

Recommended  separation  distance  between  portable  and  mobile  RF 
communications equipment and the device

This device is intended for use in an electromagnetic in which radiated RF disturbances 
are controlled. The customer or the user of this device help prevent electromagnetic 
interference by maintaining a minimum distance between portable and mobile RF 
communications equipment (transmitters) and this device as recommended below, 
according to the maximum output power of the communications equipment.

Rated maximum 

output power of 

transmitter

W

Separation distance according to frequency of transmitter

M

150 kHz to 80 MHz

d = 

80 MHz to 800 MHz

d = 

800 MHz to 2.5 GHz

d = 

0.01

N/A

0.12

0.23

0.1

N/A

0.38

0.73

1

N/A

1.2

2.3

10

N/A

3.8

7.3

100

N/A

12

23

For transmitters rated at a maximum output power not listed above, the recommended 
separation distanced in meters (m) can be estimated using the equation applicable to the 
frequency of the transmitter, where P is the maximum output power rating of the transmitter 
in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range 

applies.

NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is 

Содержание ENT001

Страница 1: ...0120 REF ENT001 WL 2103A N E U R O P E X...

Страница 2: ......

Страница 3: ...T CONTROLS ATTACHING THE LEAD WIRES ELECTRODE SELECTION AND CARE TIPS FOR SKIN CARE CONNECTING THE TENS DEVICE BATTERY INFORMATION CARING FOR YOUR TENS DEVICE TROUBLESHOOTING SYSTEM COMPONENTS TECHNIC...

Страница 4: ......

Страница 5: ...mulate the nerve or nerves in the treatment area In many cases this stimulation will greatly reduce or eliminate the pain sensation you feel by masking the original pain message sent to the brain It i...

Страница 6: ...devices for use during pregnancy or delivery has not been established TENS is not effective for pain of central origin headaches If TENS treatment becomes ineffective or unpleasant stimulation should...

Страница 7: ...Battery Strip PRECAUTIONS ADVERSE REACTIONS Isolated cases of skin irritations may occur at the site of electrode placement during long term application Effectiveness is highly dependent upon patient...

Страница 8: ...ity The intensity knobs located on the top of the unit affects the strength of the stimulation and also functions as the ON OFF control Mode The Mode switch is used to select the type of treatment uti...

Страница 9: ...kets located on top of the unit Holding the insulated portion of the connector push the plug end of the wire into one of the jacks one or two sets of the wires may be used After connecting the wires t...

Страница 10: ...when applying electrodes This is best accomplished by applying the electrode and smoothly pressing it in place from the center outward When removing electrodes always remove by pulling in the directi...

Страница 11: ...fortable reduce the stimulation amplitude to a comfortable level or cease stimulation and contact your physician if problems persist BATTERY INFORMATION If your TENS device is equipped with a recharg...

Страница 12: ...ry to force the battery Caution The battery may overheat and rupture if it is inserted backwards 2 Plug the charger and allow it to remain undisturbed for 8 to 10 hours Remove battery upon completion...

Страница 13: ...replace the battery and check again 2 If the ON OFF Indicator Light is flashing and you still feel no stimulation check that the lead wires are properly connected and the electrodes are in place If th...

Страница 14: ...ustable Cont 30 min 60 min Burst consists 2 burst per sec at 100 Hz Asymmetrical Bi Phasic square pulse 0 100 Volt open current 9 volt battery alkaline or nickel cadium rechargeable 95 H x 65 W x 23 5...

Страница 15: ...on The warranty does not apply to damage resulting from failure to follow the operating instructions accidents abuse alteration or disassembly by unauthorized individuals Batteries and electrodes are...

Страница 16: ...dance and manufacturer s declaration electromagnetic emissions below The customer or the user of this device should assure that it is used in such an environment Emissions Compliance Electromagnetic e...

Страница 17: ...5 1 kV line s and neutral Not applicable Mains power quality should be that of a typical commercial or hospital environment Voltage dips short interruptions and voltage variations on power supply inp...

Страница 18: ...to the frequency of the transmitter R e c o m m e n d e d s e p a r a t i o n distance d d 80 MHz to 800 MHz d 800 MHz to 2 5 GHz Where P is the maximum output power rating of the transmitter in watts...

Страница 19: ...turbances are controlled The customer or the user of this device help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment tra...

Страница 20: ...NEUROPEX IM 21 052 RevA17 WWW NEUROPEX CO...

Отзывы: