LKG 610
Netech Corporation
110 Toledo St, Farmingdale, NY 11735,
Controls & Indicators
Item
Name
Description
1
ECG
Connectors
Universal connectors to connect ECG snaps or various sized posts
2
Display
A 3.5 digit LCD presents the result of the selected test measurement
3
Load Selector
Switch
A slide switch on the right side of the instrument to select either the
AAMI or the IEC 601-1 test load.
4
Test Jacks
Calibrated outputs are provided for resistance (1 Ohm) and leakage
current (200
µ
A)
5
Chassis Input
Connection for the 503 test cable
6
Ground
Switch
A switch to momentarily open the ground connection to the test
receptacle
7
Polarity
Switch
A three-position switch to select the normal and reverse polarity of
the test receptacle. The center position shuts the power off to the test
receptacle.
8
Neutral
Switch
A switch to momentarily open the neutral line to the test receptacle.
9
Fuse
20 Ampere fuse for 110 VAC model, 10 Ampere fuse for 230 VAC
model
10
Serial Port
Optional RS 232 Serial Port
11
Zero
Adjustment
Recessed adjustment to zero any test cable resistance variations
12
Test
Receptacle
A 20 AMP power receptacle for the Device Under Test
13
Power Cord
The power cord supplies power to the analyzer and to the Device
Under Test
14
Keypad
Twelve push button function keys to select the test measurement
15
LEDs
An
LED in the corner of each function key indicates the selected
test mode and the unit of measurement displayed