Doc ID. 100-106 PROGRAM MANUAL.DOC
4
3. Installation
The Ness Medi is extremely easy to install and requires no tools. The normal location for the
Medi Alarm is at the first telephone installed in the premises. This ensures that the Medi Alarm
has priority if it needs to send a message to the central monitoring station. The local telephone
sits on top of the Medi Alarm unit to make it as unobtrusive as possible.
Another important aspect of the installation is the location of the Medi Alarm unit to give
optimum radio coverage. To give the best all round coverage it is best to have the Medi Alarm
installed centrally in the premises. This may not always be possible due to the location of the
telephone and mains power but it must still be a consideration. Once a suitable location is
found the install can take place.
INSTALLATION EXAMPLE
This example is based on the common Telstra T200 series telephone.
1) Select the best location, taking into consideration the customers requirements and the above
criteria.
2) Connect the plug pack and 9 volt Lithium battery and enter program mode if necessary by
pressing the ‘CANCEL’ within five seconds of powering up the system. (See programming
guide)
3) Program in any additional pendants and \ or detection devices and other program
parameters as required.
4) After which conduct a signal strength check on all devices to ensure correct radio
operation.
5) Unplug the modular telephone plug on the T200 phone (marked as ‘LINE’) and plug it into
the modular telephone socket marked ‘Line In’ on the Medi Alarm.
6) Using the supplied jumper lead with a modular telephone plug on each end. Plug one end
into the socket on the Medi Alarm marked ‘Line Out’ and the other end into the T200 phone
marked ‘LINE’.
7) Lift the handset on the telephone to check that dial tone is present. This completes the
telephone wiring process.
8) If the Medi Alarm is in program mode and no further programming is required press the
‘CANCEL’ button until ‘E’ is on the display then press the ‘HELP’ button to leave program
mode.
NOTE: It may be worth while to pre-program the system prior to arriving at the customers
premises to ensure a smooth install takes place.