5.5.2 ID control commands
#
Commands
1
:MDL?;
2
:REV?;
Description
Returns the power supply model identification as anASCII string:
Nemic-Lambda ZUP(XXV)-(YYA).
XX - The rated output voltage
YY - The rated output current
example: Nemic-Lambda ZUP(6V-33A)
Returns the software version as anASCII string:
Ver XX-YYA.B
XX- The rated output voltage
YY- The rated output current
A.B- Version identifier
example: Ver 6-33 1.0
5.5.1 Initialization control
#
Commands
2
:DCL;
1 :ADRn;
3
:RMTn;
4
:RMT?;
Description
Clears the communication buffer and the following registers:
1. Operational status register
2.Alarm (fault) status register
3. Programming error register
Sets the power supply address. ADR is followed by the address
which can be 01 to 31.
Sets the power supply to local or remote mode. (This command is active
when the unit is either in Local or Remote modes).Transition from Local to
Remote mode is made via the front panel only.
:RMT0; Transition from Remote to Local mode.
:RMT1; Transition from latched Remote to non-latched Remote.
:RMT2; Latched remote: Transition back to Local mode or to non- latched
Remote can be made via the serial port (RS232/485).
At this mode, the front panel Local/Rem function is disabled. Escape from
this mode to non -latched remote mode can be made by turning the AC
ON/OFF to OFF and after approx. 10sec.to ON again.
Returns the remote/local setting. The returned data is anASCII string.
RM1 (The unit is in remote mode)
RM2 (The unit is in latched remote mode)