2  / 13 

 

Barcode for traceability purposes 

 

Serial number 

 

Authorized representative in Europe 

 

1.3  Labeling on Package 

The following label has been adhered to the Outer Package of the BX

α

-13 or BX

α

-13A to 

assist you in the proper transport and storage of the device. 

 

1.3  Safety Information 

For your safety, please comply with the following precautions. 

 

 

Do not use in oxygen rich environments, in the presence of propane, 
gasoline or other combustible gasses, or near dust and debris. Doing so may 
cause explosion. 

 

Do not leave the illumination on for prolonged periods or cover the device 
during illumination. Doing so may cause the device to reach high 
temperatures, resulting in fire, burns or device failure. 

 

Never touch liquid which may have leaked from the batteries. In case of 
contact with the eye(s), flush with clean water without rubbing, and consult a 
physician immediately. If leaked liquid gets on skin or clothing, wash with 
clean water immediately. Failure to do so may result in serious injury, 
including loss of eyesight. 

 

Do not attempt to disassemble or modify the device. Doing so may result in 
electric shock, fire or device failure. 

 

Do not touch the device with wet hands. Doing so may result in electric 
shock. 

 

Do not touch the Lightbulb immediately after turning the illumination off, as 
the bulb may be extremely hot. Doing so may result in burns. 

Prohibited

 

Prohibited

 

Prohibited

 

Prohibited

 

Prohibited

 

Prohibited

 

Environmental conditions for 

transport and storage

 

Содержание BXa-13

Страница 1: ...ion of the internal areas and fundus of the eye This device is available in 2 models depending on the size of the batteries used Ophthalmoscope BX 13 hereinafter referred to as BX 13 using 2 C size ba...

Страница 2: ...Turning Illumination ON OFF 5 3 2 Correction Lens and Filters 5 3 2 1 Rekoss Lens 5 3 2 2 4000 K and Polarizing Filters 5 3 2 3 Observation Polarizing Filter 6 3 2 4 Illumination Dial 6 3 3 Viewing th...

Страница 3: ...ou of potential hazards to yourself and others Potentially hazardous situation which if not avoided could result in death or serious injury Potentially hazardous situation which if not avoided may res...

Страница 4: ...reach high temperatures resulting in fire burns or device failure Never touch liquid which may have leaked from the batteries In case of contact with the eye s flush with clean water without rubbing a...

Страница 5: ...ng so may cause device failure Do not use the BX 13 or BX 13A adjacent to or stacked with other equipment In cases where adjacent or stacked use is necessary verify the BX 13 or BX 13A operates normal...

Страница 6: ...ry Lens Selector Diopter Indicator Illumination Dial Filter Lever Observation Polarizing Filter Lightbulb Manufacturer s Logo Model Logo Country of Origin CE Mark Serial Number of Head Unit 2 Battery...

Страница 7: ...In the above example it would be necessary to set the Rekoss Disc to 4 D However in practice the Rekoss Disc is rotated while observing the fundus to find the appropriate diopter In cases of extreme...

Страница 8: ...nute fundal hemorrhages Concentric Scale While observing the patient s fundus instruct the patient to fixate on the center of the concentric scale It is possible the determine the presence of eccentri...

Страница 9: ...of this red reflex slowly move closer with the ophthalmoscope to the patient s face The fundus image should become visible at a distance of about 2 cm to 3 cm see Fig 3 3 9 Rotate the Rekoss Disc as...

Страница 10: ...ay be extremely hot Doing so may result in burns If the BX 13 or BX 13A does not light the Lightbulb may have blown out Follow the directions below to check if the Lightbulb has blown out and if so to...

Страница 11: ...Illumination blocked Diopter Indicator lights up Illumination Dial mispositioned Rotate the Illumination Dial until it stops with a click Filter Lever mispositioned Swing the Filter Lever until it sto...

Страница 12: ...owered ME equipment Applied parts No applied parts Degree of protection against harmful ingress of water or particulate matter IPX0 Method s of sterilization Do not sterilize Suitability for use in an...

Страница 13: ...located at the following address LOCAL DEALER Neitz Instruments Co Ltd 4F Ichibancho Court 15 21 Ichibancho Chiyoda ku Tokyo 102 0082 Japan Phone 81 3 3237 0552 Fax 81 3 3237 0554 e mail int neitz co...

Страница 14: ...IEC 60601 1 2 2014 a Connection to the public mains network None b Applicable port s Enclosure Port Electromagnetic Emissions Phenomenon Applied Standard Emission Test Levels Conducted and radiated R...

Страница 15: ...13 13 Notes...

Страница 16: ...Japan Phone 81 3 3237 0552 FAX 81 3 3237 0554 E mail int neitz co jp URL http www neitz ophthalmic com Distributed by As part of our policy of continued product improvement we reserve the right to alt...

Отзывы: