4  Maintenance 

4.1  Cleaning 

If the device becomes dirty, wipe clean with a dry, lint-free cloth. 
Should dirt remain, moisten with diluted neutral detergent and wipe clean. Then, wipe dry. 
To disinfect the device, only wipe with alcohol, and DO NOT STERILIZE. 
Use an air blower or the like to remove dust and debris from the Viewing Window. 

4.2  Replacing Batteries 

As the batteries become depleted, the illumination will become dark, or will not light at all. 
In such cases, follow the directions below to replace the batteries.   

(1)  Remove the Bottom Cap from the 

Handle by turning counter-clockwise. 

(2)  Remove both batteries. 
(3)  Insert two (2) new batteries with the plus 

terminal going in first (see Fig. 4.1). 

(4)  Screw the Bottom Cap shut by turning 

clockwise. 

4.3  Replacing the Lightbulb 

 

WARNING 

Do not touch the Lightbulb immediately after turning the illumination off, as the 
bulb may be extremely hot. Doing so may result in burns. 

If the BX

α

-13 or BX

α

-13A does not light, the Lightbulb may have blown out. Follow the 

directions below to check if the Lightbulb has blown out, and if so, to replace the Lightbulb. 

(1)  Holding the Head in one hand and the Handle in 

the other, turn the Handle counter-clockwise to 
detach from the Head (see Fig. 4.2).   

(2)  Remove the Lightbulb from the Head by pulling the 

tip which can be found at the bottom of the Head. 

(3)  Check the Lightbulb, and if the filament has blown 

out, replace with a new bulb. 

(4)  Insert the new Lightbulb until it comes to a stop. 

When inserting the bulb, align the protrusion on the 
bulb with either of the grooves located at the 
bottom of the Head (see Fig. 4.3). Pay close 
attention so as not to touch the glass part of the 
bulb. 

(5)  Attach the Head and Handle by screwing the 

Handle clockwise. 

4.4  Disposal 

Dispose of the BX

α

-13 or the BX

α

-13A and its accessories in accordance with your local 

regulations and/or environmental guidelines. 
 

 

Fig. 4.1 

Handle 

Bottom Cap 

Direction of Batteries 

Fig. 4.2 

Fig. 4.3 

Protrusion 

Содержание BXa-13

Страница 1: ...ion of the internal areas and fundus of the eye This device is available in 2 models depending on the size of the batteries used Ophthalmoscope BX 13 hereinafter referred to as BX 13 using 2 C size ba...

Страница 2: ...Turning Illumination ON OFF 5 3 2 Correction Lens and Filters 5 3 2 1 Rekoss Lens 5 3 2 2 4000 K and Polarizing Filters 5 3 2 3 Observation Polarizing Filter 6 3 2 4 Illumination Dial 6 3 3 Viewing th...

Страница 3: ...ou of potential hazards to yourself and others Potentially hazardous situation which if not avoided could result in death or serious injury Potentially hazardous situation which if not avoided may res...

Страница 4: ...reach high temperatures resulting in fire burns or device failure Never touch liquid which may have leaked from the batteries In case of contact with the eye s flush with clean water without rubbing a...

Страница 5: ...ng so may cause device failure Do not use the BX 13 or BX 13A adjacent to or stacked with other equipment In cases where adjacent or stacked use is necessary verify the BX 13 or BX 13A operates normal...

Страница 6: ...ry Lens Selector Diopter Indicator Illumination Dial Filter Lever Observation Polarizing Filter Lightbulb Manufacturer s Logo Model Logo Country of Origin CE Mark Serial Number of Head Unit 2 Battery...

Страница 7: ...In the above example it would be necessary to set the Rekoss Disc to 4 D However in practice the Rekoss Disc is rotated while observing the fundus to find the appropriate diopter In cases of extreme...

Страница 8: ...nute fundal hemorrhages Concentric Scale While observing the patient s fundus instruct the patient to fixate on the center of the concentric scale It is possible the determine the presence of eccentri...

Страница 9: ...of this red reflex slowly move closer with the ophthalmoscope to the patient s face The fundus image should become visible at a distance of about 2 cm to 3 cm see Fig 3 3 9 Rotate the Rekoss Disc as...

Страница 10: ...ay be extremely hot Doing so may result in burns If the BX 13 or BX 13A does not light the Lightbulb may have blown out Follow the directions below to check if the Lightbulb has blown out and if so to...

Страница 11: ...Illumination blocked Diopter Indicator lights up Illumination Dial mispositioned Rotate the Illumination Dial until it stops with a click Filter Lever mispositioned Swing the Filter Lever until it sto...

Страница 12: ...owered ME equipment Applied parts No applied parts Degree of protection against harmful ingress of water or particulate matter IPX0 Method s of sterilization Do not sterilize Suitability for use in an...

Страница 13: ...located at the following address LOCAL DEALER Neitz Instruments Co Ltd 4F Ichibancho Court 15 21 Ichibancho Chiyoda ku Tokyo 102 0082 Japan Phone 81 3 3237 0552 Fax 81 3 3237 0554 e mail int neitz co...

Страница 14: ...IEC 60601 1 2 2014 a Connection to the public mains network None b Applicable port s Enclosure Port Electromagnetic Emissions Phenomenon Applied Standard Emission Test Levels Conducted and radiated R...

Страница 15: ...13 13 Notes...

Страница 16: ...Japan Phone 81 3 3237 0552 FAX 81 3 3237 0554 E mail int neitz co jp URL http www neitz ophthalmic com Distributed by As part of our policy of continued product improvement we reserve the right to alt...

Отзывы: