NED
XCM8060SA_8040SA_6040SA
UME-0034-03
56
(3) Capture images using a camera interface board utility. In the case of Matrox’s
Solios, it is convenient to use Intellicam.
Figure 5-2-3 Solios Window
5.3 During Operation
(1) Does an acquisition time out error occur?
<Cause>
<1> Captured images are too large.
If there are many filtering processes, the assignments to the driver may be
insufficient.
<2> The cable is detached from the connector
Ensure that the power cable and Camera Link cables are connected to the
camera firmly.
<3> Camera Link cables are susceptible to noise when the cables are laid near a
light source inverter line or a power line. The personal computer in use may be
freeze and need to be reset.
(2) Are there dark lines in the direction of vertical scanning on the image?
<Cause>
<1> Dust on the sensor window
Dust may get onto the sensor window from the inside or the outside of the camera.
Remove the dust with air or a lens cleaner.