English-15
English
Additional Installation Information
Advanced OSD menu
Several control functions are available by entering the advanced OSD menu. These control functions are additional OSD menu
items, which are always accessible after opening the standard OSD menu.
Control Function
Description
EDID
Change the picture format to be recognized by the operating system. The factory default is
“(Auto Selection)”.
•
Portrait format
- MD212G3: 1536 x 2048 pixels.
- MD211G5: 2048 x 2560 pixels.
•
Landscape format
- MD212G3: 2048 x 1536 pixels.
- MD211G5: 2560 x 2048 pixels.
Display Mode
(MD211G5 only)
Change the video signal input format
•
Normal Mode
[Factory Default]
The display accepts the video signal with either 25Hz refresh rate (DVI Single Link) or 50Hz
refresh rate (DVI Dual Link), based on DVI 1.0 specifi cations.
•
Packed Pixel Mode
The display accepts the video signal with 50Hz refresh rate in a special packed pixel data format
(DVI Single Link). This requires the use of a customized driver for the display controller. Make
sure that a customized driver is available for your specifi c display controller model. For the
availability of a packed pixel driver, please contact your supplier.
Input Source
Change the input source
•
DVI
•
DisplayPort
Confi guration
Change luminance and gamma correction factory presets. The factory presets are:
MD212G3:
Confi guration
Luminance (cd/m
2
)
Gamma
Remark
Maximum
Minimum
Ambient
1
400
0.6
0.0
DICOM
Default
2
400
1.0
0.0
DICOM
3
500
0.6
0.0
DICOM
MD211G5:
Confi guration
Luminance (cd/m
2
)
Gamma
Remark
Maximum
Minimum
Ambient
1
500
1.0
0.0
DICOM
Default
2
550
1.0
0.0
DICOM
3
400
1.0
0.0
DICOM
The confi guration 1 parameter may be modifi ed anytime by a new display calibration.
NOTE:
Regarding display calibration, please refer to the GammaCompMD QA user manual.
DICOM
conformance test
Perform a quick DICOM conformance test without software to check the display properties.
Содержание MD211G5
Страница 1: ...MD212G3 MD211G5 ...