6.3.3 Average times settings
It is extremely necessary to work out the average measuring values in leeb's
hardness testing. Different measuring times between 2~8 are selected
according to users’ different requirements.
- Press the
key and the
key to enter
MEASURE-
AVERAGE
.
- Press the
key to enter this item, use the
key or
key
to set the desired average times.
- Press the
key to confirm.
Prompt: 3~5 times are usually recommended in normal measurement.
6.4 Files management
Via “file management”, users are allowed to set the file management ways.
For example, users can set the saved file codes according to their own
requirements.
6.4.1 New files settings
Via this sub-menu, users are allowed to set the number of files freely.
DHT-200PLUS hardness tester supplies users 100 files, which are coded from
F001 to F100. Per each file, there is also 100 data storage, which is coded
from 001 to 100. For example it shows F001-099 on LCD display, it means the
99
th
test data in File No. 001.
For the convenience of saving the data for different work pieces, the
function of work piece marking is available in the current file. Then users will
be able to identify the data of different work pieces saved in the same file.
Users are also allowed to build only one file. When the auto-save is turned
on, the indication mark
appears on the screen, and the measured values
will be saved in the file in a series of codes. When the memory is full, it will
give a prompt and said that “please build another file” or “the memory is
over”, “please delete the data unused”. When a new file is built, the former
one will be closed automatically, and at the same time it is featured “read
only”, which means that users can only read or print out the former file but
not write in.
Keys
and
for setting
,
key
for
confirmation