background image

10

10

IP SUPPRESSION

 (IP-SUP)

Frequently transducers exhibit a broad roll off at the Initial Pulse. The IP Suppression can
frequently compensate for this thereby permitting measurements closer the initial pulse when in
Contact Mode. In many cases, a fine tuned setup will permit readings as thin as 0.020" using a 5
MHz 1/4" Transducer.

MEASUREMENT TIME

 (MEAS-TIM)

Measurement Time is a unique feature generally used in Delay Line
Transducer Measurements. In the figure to the left we have used a
standard NDT Systems, D11 Transducer. Currently the gage is
triggering on the 2

nd

 delay line multiple which is of no value and could

potentially cause operator confusion when actual measurements
approach the delay end. Therefore, the MEAS-TIM control acts like a
blocking gate except but acts on the tail end of the measured zone.
The area we highlighted in the figure is an area to block out. When
this is done, any time the probe is coupled it will read over range
rather than a fixed reading to the delay end.

DISPLAY MENU

FREEZE

When activated stops any screen activity and “holds” whatever was
on the screen at the time the freeze option was selected. Activate the
function by using the ENTER key.

CONTRAST

 (CNTRST)

Using the up and down arrow keys, adjust the display to provide
optimal personal viewing.

BACKLIGHT

 (BKLT)

Turns the backlight to AUTO, ON or OFF. Auto will turn the backlight

on while making measurements. OO will turn the backlight on continuously
and OFF will leave the backlight off at all time. It is recommended to leave
the setting set to off if using the gage in high ambient light conditions. This
will conserve battery life. You will not notice any better viewing while
having the backlight on in high ambient light conditions.

The Following Features are NEW and Unique Productivity Features.

Note the figure to the left. You will note there are two traces annotated ‘A’
& ‘B’. The top trace ‘A’ is called the MAIN Trace Window. The Lower
Trace, ‘B’ is known as the Tracking Window. Also, take note of the Vertical
Cursors. The area between these Cursors is the section of the echo which
will be automatically Zoomed as well as Tracked. Meaning, as the echo
moves left and right on the Main Trace, the actual area of the echo is
Tracked to show the half cycle the gage is triggering on.

SIZE

Pressing the up and down arrow key will resize the ratio of the Main
Trace Window to the Tracking window. Pressing the Up Arrow in-
creases the Main Window size until there would only be one A-Trace
Screen presented, representing the Main Trace Window with no
Tracking window. Conversely, if you press the Down Arrow Key, the
Tracking window will increase in size, relative to the Main Trace. In this case the Track window

Meas-TIM Zone

MEAS-TIM Set

Содержание TG400

Страница 1: ...ERATOR S MANUAL TG400 Ultrasonic Thickness Gage Featuring SplitView SplitScan AutoTrack NDT Systems Inc 17811 Georgetown Lane Huntington Beach CA 92647 Phone 714 893 2438 FAX 714 897 3840 Rev 1 0 07 0...

Страница 2: ...may have a question with C Chapter 2 deals with the menu structure a description of the features This would be recommended reading There are a few images to help you better understand the concepts C C...

Страница 3: ...0 000 nanoseconds Synchronization IP Initial Pulse IF Interface Delay Transducer Modes Single or dual element Display Modes RF positive negative full wave rectified Flaw Mode Two 2 gates selectable po...

Страница 4: ...ypical depending upon display back light usage 3 AA NiMH furnished Battery Charger 5 hours for full charge Transducer Cable Lemo Connectors Connectors Size Top Section 3 25 W x 7 L x 1 4 D Weight Incl...

Страница 5: ...ety of keys menu selections are available to change continuous variables such as range delay gate positions alarm levels and reject SCROLLING When a key is continuously depressed to change a variable...

Страница 6: ...A free catalog will be sent on request by contacting Customer Information at NDT Systems Inc 17188 Georgetown Lane Huntington Beach CA 92647 phone 714 893 2438 FAX 714 897 3840 In many cases equivale...

Страница 7: ...plit Opt1 is reserved for future use MENU FLOW The following represents the menu flow of the TG400 Pressing the MENU ESC key repeatedly will sequence through the menus as shown Note The branch if menu...

Страница 8: ...using many kinds of transducers to be successfully performed highly precise flaw detection tests may require refinements to the default variables to better match transducer instrument characteristics...

Страница 9: ...useful ness of this feature would be evident on material with varied attenuation properties such as painted and non painted surfaces or corroded back surfaces AGC may also help overall reading accura...

Страница 10: ...rease or Decrease the Measurement Threshold level Any echo or echo half cycle with amplitude equal to or greater than the Thickness Gate Threshold will be measured The threshold gate is the bar on the...

Страница 11: ...vide optimal personal viewing BACKLIGHT BKLT Turns the backlight to AUTO ON or OFF Auto will turn the backlight on while making measurements OO will turn the backlight on continuously and OFF will lea...

Страница 12: ...can not exceed the set Range of the Main Trace window For instance if the Main Trace is set to 3 the Track window can not be set to 4 Usually the Track window is something like 20 40 of the Main Trac...

Страница 13: ...12 12 instrument setups Note If the gage is powered ON with the ENTER button depressed the Factory Default setup will be loaded...

Страница 14: ...at the water path to the IF will be seen as well as the IF to destination target back wall Be sure this rule stays in place throughout the water path change probability It is OK to select a long MAIN...

Страница 15: ...es However such transducers will be relatively expensive and not readily attainable Standard inexpensive delay line transducers highly damped in the frequency range of 10 to 20 MHZ will ordinarily pro...

Страница 16: ...nd be used over a broad range of thicknesses test material attenuation may dictate a compromise While higher frequencies produce thinner thickness resolution attenuation is more pronounced at higher f...

Страница 17: ...llic Test Materials Virtually all the techniques and applications described above can be applied to a host of non metallic and composite materials It is not possible to cover the numerous applications...

Страница 18: ...way of observing these effects and to become quickly familiar with TG400 s controls is to perform the following procedure 1 Attach a highly damped broadband transducer such as the OPTIMA CHG053 or NO...

Страница 19: ...ive gain accordingly 11 To become familiar with the RANGE function position the cursor at RNG The up arrow key will squeeze the echos on the A trace to the left that is except for the initial pulse If...

Страница 20: ...t this point use the Up Dn arrow keys to change the length of the IP BLK gate in order to observe the effects of the control 16 The IF BLK interface echo blocking functions in a similar fashion as the...

Страница 21: ...Thickness readout the upper range calibration is complete Note that zero may change through up to 5 or 10 counts on the display before the indicated thickness changes Choosing zero roughly at the mid...

Страница 22: ...when using single element transducers with thin elastomeric membranes or on coated test objects 23 Cursor to ECHO BLOCKING GATE ECHO BLK Note the thick horizontal bar on the Thickness gate Use the rig...

Страница 23: ...ponents This mode of display RF reveals all the details of the signal If necessary to more faithfully reproduce the waveform in the figure above it may be necessary to key to GAIN and change the gain...

Страница 24: ...ange of the TG400 has been set to a long range and the RF display is selected there will be poor detail in the presentation of the RF In order to expand the RF display of the back echo under such cond...

Страница 25: ...he constant distance paths taken by the sound beam in traversing the dual delay lines a measurement can be made of the travel distance in the test object only Since the sound beam travels at a slight...

Страница 26: ...O to produce the thickness readout corresponding to a known thickness in this case 0 500 inch Note when you check the measurement of the 0 200 inch step note it produces 0 205 inch This error is due t...

Страница 27: ...0 If it close adjust the VEL to adjust to exact material velocity Option B Connect TG506 Transducer From MAIN Menu Select SETUP and PICK the appropriate transducer from the list Press ZERO ECHO front...

Страница 28: ...n this case the ECHO BLK will need to be extended out far enough to pass the multiple echo cycles of the first echo This will limit the low end thickness capability In the lower image a TG560P transdu...

Страница 29: ...ve right when you move the cart to the left the B Scan Cursor move to the right ENCODED L L Left to Left When you move the cart to the right the B Scan Cursor move right when you move the cart to the...

Страница 30: ...on screen from the thickest area to the thinnest area of your sample 2 Go to the SETUP Menu and select the PX506 if it is shown If not select TG506 This should set the instrument up in general terms A...

Страница 31: ...ld not have to go through the extra steps 11 Once calibrated press the B SCAN Key 12 Cursor to TYPE and use the Up Dn arrows to select T SCAN This will provide a Cross Section view of the area being i...

Страница 32: ...racteristics such as the presence of scattering reflectors impedance an isotropy Modula and other physical and mechanical properties all interact with ultrasonic sound beams somewhat differently than...

Страница 33: ...AA style batteries inserted into the bottom panel via the rear panel access cap The gage is charged via an external battery charger Once fully charged the TG400 will run for up to 20 hours before req...

Страница 34: ...ng optional TG400 accessories are available ATRP1 TG400 Carry Pouch A light weight leatherette protective case The TG400 fits firmly into a case like compartment supported at an adjustable angle stand...

Страница 35: ...exclusive remedy provided herein fails of its essential purpose or operates to deprive either party of the substantial value of its bargain then the Purchaser s exclusive remedy will be the return of...

Отзывы: