Manual No. OM5000
Jan X/04 Ver 1.40
41
TABLE III
Thickness Gaging Limits During Immersion
Or Delay Line Transducer Gaging
RANGE SWITCH
POSITION
SYNC T-GATE
SWITCH POSITION
NOMINAL
ECHO PERIOD (us)
EQUIV.STEEL
THICK
1"
1st
6.40
0.75"
1"
2nd
3.10
0.36"
1"
3rd
2.20
0.25"
10"
1st
6.41
4"
10"
2nd
3.19
2"
10"
3rd
2.26
1.3"
10mm
1st
6.4
19.1mm
10mm
2nd
3.1
9.1mm
10mm
3rd
2.2
6.4mm
100mm
1st
34.3
101.6mm
100mm
2nd
17.2
50.8mm
100mm
3rd
11.2
33.0mm
TABLE IV
Waterpath Limits During
Immersion Transducer Gaging
RANGE SWITCH
NOMINAL WATERPATH LIMITS*
POSITION
MINIMUM
MAXIMUM
1"
0.2"
Rep Rate Dependent
10"
1.0"
Rep Rate Dependent
10mm
5.1mm
Rep Rate Dependent
100mm
25.4mm
Rep Rate Dependent
*Rep Rate Overlap (see Section 6.3) and Water Multiples Must Be
Considered with Above Information.