Manual No. OM5000
Jan X/04 Ver 1.40
25
NovaScope.
NOTE: the optional BUDdy card must be installed prior to the NovaScope being
turned on. If it is not, save the existing setup internally and exit to the front panel view and turn
the unit off. Insert the BUDdy card into the rear panel slot and then restart the 5000.
The following
is an example procedure for the saving of a setup from one NovaScope for recall to another
NovaScope.
a)
From the front panel select setup
b)
If this is new setup select the NEW button and save the setting “locally” or internally using the
pop-up keypad and then select ENTER on the keypad. You will be returned to the Save-
Recall screen. Notice the text above the keyboard. At this point is is reminding you that the
setting will be save internally.
c)
Press the EXT button. You should notice that the text over the Save-Recall screen reflects
External device selected.
d)
You will now notice the EXT button name has changed to INT (internal). You may now select
STORE to save the existing setup to the EXTernal device under the same name as you did
under the INTernal setup name. Otherwise, you may select NEW and the Keyboard will be
displayed permitting you to save the setup under a new name to the EXTernal BUDdy card.
e)
If you are interested on moving this setup to another NovaScope, Select the INT button and
then exit to the front panel and turn NovaScope unit one off. Remove the BUDdy card from
unit one and insert it to unit two. Turn on unit two.
f)
Select SETUP, EXT
g)
Select the desired EXT file name
h)
Press RECALL
i)
Press the INT button
j)
Press STORE and the file will be saved internally on unit two as the same name.