Introduction
Operator’s Manual V1.1 • EasyOne Air • EasyOne Connect
© ndd Medizintechnik AG • Technoparkstrasse 1, 8005 Zurich, Switzerland • www.ndd.ch
39
When
EasyOne Air
is turned on and not in use, the screensaver and automatic
power-off function are activated to save battery power. You can change the
default configuration in the following menu:
>
>
> Screensaver after
and
Power-off after
A low battery message will alert you when battery power falls below 10%.
Data saved in non-volatile memory is not lost when battery power is low or
when batteries are removed.
The
EasyOne Flow Tube
respiratory tube for single-patient use
The
EasyOne Flow Tube
respiratory tube
To ensure hygienic testing, the
EasyOne Flow Tube
respiratory tube is a
disposable for single-patient use.
EasyOne Air
recognizes the correct position
of the
EasyOne Flow Tube
in the
EasyOne Flow Tube
holder and gives feedback
if the
EasyOne Flow Tube
is not placed correctly. The
EasyOne Flow Tube
is
designed to work accurately with the ultrasonic sensor of
EasyOne Air
. For
reliable results, use only the original
ndd
EasyOne Flow Tube
.
CAUTION
FALSE DIAGNOSIS
Non-original respiratory tubes can cause measurement error and false results.
You must only use
EasyOne Flow Tube
respiratory tubes by the manufacturer
ndd
to assure accuracy, long-life, and full warranty coverage.