Maximum charging method
1) 0.7C constant current up to 4.2 V
2) Constant voltage of 4.2 V until charge current declines to
≤
0.01C
Charge cut-off voltage
4.2 V to 4.25 V
Standard discharge current (C-rate)
0.2C constant current down to end voltage of 3.0 V
Maximum discharge current
(C-rate)
0.7C constant current down to end voltage of 3.0 V
Discharge cut off- voltage
2.8 V to 3.0 V
Initial impedance
≤
200 m
Ω
The following standard conditions are identical to the conditions of
EasyOne Air
. Storage and operation under these
conditions is safe.
Standard storage conditions
Temperature
-20°C to 50°C (-4°F to 122°F)
Relative humidity
5% to 90%
Atmospheric pressure 500 hPa to 1060 hPa
Battery
removed battery
Standard operating conditions
Temperature
0°C to 40°C (32°F to 104°F)
Relative humidity
5% to 90%
Atmospheric pressure 700 hPa to 1060 hPa
The following optimized conditions prolong the life expectancy of the battery.
Optimized environmental
conditions
Temperature
23±5°C (73.4±9°F)
Humidity
45% to 75% relative humidity
Atmospheric Pressure 860 hPa to 1060 hPa
Optimized operating conditions
while charging and discharging
Charging
0°C to 45°C (32°F to 113°F)
45% to 85% relative humidity
Discharging
-20°C to 60°C (-4°F to 140°F)
45% to 85% relative humidity
Optimized storage periods and
storage conditions resulting in no
less than 80% recoverable capacity
of the initial capacity
Storage period
≤
1 month
-20°C to 60°C (-4°F to 140°F)
45% to 85% relative humidity
Storage period
≤
3 months
-20°C to 45°C (-4°F to 113°F)
45% to 85% relative humidity
Storage period
≤
1 year -20°C to 20°C (-4°F to 68°F)
45% to 85% relative humidity
Specifications for the lithium-ion battery of