Universal XactTrace
®
User Instructions
13
Table 4
– Immunity Test Levels – Enclosure Port
Phenomenon
Basic EMC
Standard or Test
Method
Immunity Test Levels
– Professional
Healthcare Facility Environment
Electrostatic Discharge
IEC 61000-4-2
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV, ± 15 kV air
Radiated RF EM Fields
IEC 61000-4-3
3 V/m
80 MHz
– 2,7 GHz
80 % AM at 1 kHz
Proximity Fields from
RF Wireless
Communications
Equipment
IEC 61000-4-3
See “Enclosure Port Immunity to RF
Wireless Communications Equipment”
Table below
Rated Power
Frequency Magnetic
Fields
IEC 61000-4-8
30 A/m
50 Hz and 60 Hz
Table 8
– Immunity Test Levels – Signal Input / Output Parts
Port
Phenomenon
Basic EMC
Standard
Immunity Test Levels
–
Professional Healthcare Facility
Environment
Conducted
Disturbances
Induced by RF
Fields
IEC 61000-4-6
3 V
0,15 MHz
– 80 MHz
6 V in ISM bands between
0,15 MHz and 80 MHz
80 % AM at 1 kHz