background image

Quick Guide: 

Bio-logic

®

 AuDX PRO FLEX

®

 Diagnostic OAE and audiometry

Otometrics - Next generation Bio-logic products

9

Bio-logic

®

Connect the red probe to the red 
connector and select one of the 
Presets for DPOAE Threshold. To 
perform binaural testing, also 
connect the blue probe to the 
blue connector.

Select the appropriate ear tip size 
and place it on the probe tip. 
Insert the probe in the ear making 
sure that the ear tip fits tightly into 
the ear canal.

Select the test ear by pressing 
either the 

Right

 or 

Left

 button, 

or the 

Right + Left 

button for 

binaural testing.

Cartoon mode will start the test 
showing a cartoon instead of the 
progress bar, this can be changed 
during the test by tapping on the
plane icon at the bottom of the 
device screen.

DPOAE threshold

A calibration will be performed prior to measurement. A green light indicates successful calibration.
After successful ear probe calibration, the test will start automatically.

Probe fit override for patients with
PE tubes or middle ear perforations

Skip a frequency

Pause/resume test

Stop test

Progress for each 
tested frequency 
in each ear

cartoon mode 
for children

Noise floor level

Содержание Otometrics Bio-logic AuDX PRO FLEX

Страница 1: ...5 55 55 otoinfo natus com Natus Medical Incorporated 1 800 289 2150 otoorder us natus com www otometrics com Bio logic AuDX PRO FLEX Diagnostic OAE and audiometry Quick start guide Doc No 7 26 1375 EN 00 Part No 7 26 13705 EN Bio logic ...

Страница 2: ... off button on right side Current time Battery status Select patient Additional information on options for current screen Turn off device Customize settings Power supply connector Connect the headphones to the blue connector Select Audio Threshold preset from the start menu Select the ear on which you want to start the test The test screen will be displayed Testing Audiometry ...

Страница 3: ... indicator Select the ear to test Select the stimulus levels Additional information about options for current screen Change settings Enter masking mode Stop test Select the frequency Test screen Audiometry test results Green background indicates range of normal hearing Print results to a label or to a pdf file or full page report on PC ...

Страница 4: ...t the headphones to the blue connector Select Audiometry screening Preset from the start menu Select the ear on which you want to start the test The test screen will be displayed Select the stimulus levels Select the type of stimulus presentation Select the ear to test Select the frequency Press to present the stimulus Stop test Additional information about options for current screen ...

Страница 5: ...of the Presets for DPOAE Quick To perform binaural testing also connect the blue probe to the blue connector Select the appropriate ear tip size and place it on the probe tip Insert the probe in the ear making sure that the ear tip fits tightly into the ear canal Select the test ear by pressing either the Right or Left button or the Right Left button for binaural testing Cartoon mode will start th...

Страница 6: ...ar after both ears have been tested Print results to a label or to a pdf file Note These circles indicate that other screens are available to view Swipe your finger right or left to get other views Retest frequencies that did not pass Test other ear Green check marks indicate valid response Red circles indicate no valid response DPOAE screening test results Please note that the initial result scre...

Страница 7: ...ppropriate ear tip size and place it on the probe tip Insert the probe in the ear making sure that the ear tip fits tightly into the ear canal Select the test ear by pressing either the Right or Left button or the Right Left button for binaural testing Cartoon mode will start the test showing a cartoon instead of the progress bar this can be changed during the test by tapping on the plane icon at ...

Страница 8: ...left to get other views Print results to a label or to a pdf file Test other ear Green check marks indicate valid sponse Red or blue bars indicate response grey bars indicate noise High resolution test results 4 PPO Outcome for each test sorted by f2 and L2 Multilevel test results shown Tap on screen to view details for each level tested Numeric values shown in table format DPOAE diagnostic high r...

Страница 9: ...ar by pressing either the Right or Left button or the Right Left button for binaural testing Cartoon mode will start the test showing a cartoon instead of the progress bar this can be changed during the test by tapping on the plane icon at the bottom of the device screen DPOAE threshold A calibration will be performed prior to measurement A green light indicates successful calibration After succes...

Страница 10: ...AuDX PRO FLEX Diagnostic OAE and audiometry Otometrics Next generation Bio logic products 10 Bio logic Green background indicates range of normal hearing Print results to a label or to a pdf file DPOAE threshold test results ...

Страница 11: ...rves the right to make changes at any time without notice Please note that not all products displayed on our webpage or advertising materials are available in all territories Please contact your local distributor or PATH medical for further information The latest revision of the Quick Start Guide is available online at www pathme de download Errors and omissions excepted Copyright Notice No part o...

Отзывы: