Testing the Warmette
Testing the Warmette
Test the Warmette after installing it or performing any maintenance or service
procedure to ensure that all systems are functioning properly.
To prepare the Warmette for testing:
1
Plug in the power cord, then press the power switch to the
ON
(|) position.
2
Set the temperature to 120°F.
3
Allow the Warmette to warm up for 5 minutes.
To test the Warmette:
Ensure that the
HEATER ON
lamp is illuminated.
Ensure that the air blowing from the top of the Warmette is noticeably warm.
Listen for the blower to make sure it is running.
Check that the
HEATER ON
lamp turns off within 60 minutes. It will then cycle
on and off to maintain the set temperature.
Electrical Requirements
The Warmette product line has several voltage/frequency options:
120V~, 60 Hz, 14A
220V~, 50 Hz, 8A
220V~, 60 Hz, 8A
CAUTION
This product must be connected directly to a hospital-grade electrical outlet(s) rated for
minimum 20-ampere (A) service. Each heating cabinet is electrically separate. Two
separate electrical outlets are required for the Model 55/56.
Содержание Olympic Warmette 48
Страница 2: ......
Страница 4: ......
Страница 8: ...iv Olympic Warmette Models 48 49 and 55 56 Instruction Manual ...
Страница 10: ...vi Olympic Warmette Models 48 49 and 55 56 Instruction Manual ...
Страница 14: ...1 4 Olympic Warmette Models 48 49 and 55 56 Instruction Manual Description ...
Страница 20: ...2 6 Olympic Warmette Models 48 49 and 55 56 Instruction Manual Electrical Requirements ...
Страница 28: ...4 6 Olympic Warmette Models 48 49 and 55 56 Instruction Manual Annual Maintenance ...
Страница 44: ...7 2 Olympic Warmette Models 48 49 and 55 56 Instruction Manual Ordering Information ...