Revised 11/21/18
4-7
Electromagnetic Compatibility (EMC)
Table 8 – Immunity Test Levels - Signal Input / Output Parts Port
Phenomenon
Basic EMC Standard
Immunity Test Levels - Professional Healthcare
Facility Environment
Electrostatic Discharge
IEC 61000-4-2
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV, ± 15 kV air
Electrical Fast Transients/Bursts
IEC 61000-4-4
± 1 kV
100 kHz repetition frequency
Surges Line-to-Ground
(Common Mode)
IEC 61000-4-5
± 2 kV
Conducted Disturbances Induced
by RF Fields
IEC 61000-4-6
3 V
0,15 MHz – 80 MHz
6 V in ISM bands between 0,15 MHz and 80 MHz
80% AM at 1 kHz