Revised 11/21/18
3-13
Performing a Cortical Stimulation Study
IMPORTANT: The stimulus output is stopped automatically if the current
delivered exceeds the requested current level by 20% or greater. The Nicolet
Cortical Stimulator must be turned off and then turned back on to clear this
condition.
If you are currently stimulating the patient, press the Stop button on the
Stimulus Control Unit (SCU) before changing any parameters.
The following parameters can also be set through the Nicolet system PC when
the Stimulus Control Unit (SCU) and PC are connected via the RS232 port.
1.
Setup
- Press the
Setup
button;
Change Mode
appears on the display and
begins to flash.
2.
Change Mode
- Press the
Selector
knob to select
Change Mode
. You can
choose from two options in the Change Mode function:
Change Mode
Probe Biphasic
Electrode Biphasic
3.
Selector Knob
- Rotate the
Selector
knob to choose
Probe
Biphasic
stimulus if using the probe or
Electrode Biphasic
if using the Stimulus
Switching Unit (SSU). Press the
Selector
knob to make your selection.
4.
Other Functions
- To select any of the other four Setup functions (see
below), rotate the
Selector
knob and then press the
Selector
knob when the
desired function is displayed. Rotate the
Selector
knob until the desired
parameter is displayed and then press the
Selector
knob to make your
selection.
Select Label Set
Numeric
Montage
Delete Label Set
Select Language
A list of available languages appears
Adjust Contrast
Rotate the Selector knob for the best display contrast
Содержание Nicolet
Страница 2: ......
Страница 6: ...Nicolet Cortical Stimulator 4 Revised 11 21 18 Blank page ...
Страница 11: ...Revised 11 21 18 1 1 1 General Information ...
Страница 12: ...Nicolet Cortical Stimulator 1 2 Revised 11 21 18 Blank Page ...
Страница 23: ...Revised 11 21 18 2 1 2 Nicolet Cortical Stimulator Overview ...
Страница 24: ...Nicolet Cortical Stimulator 2 2 Revised 11 21 18 Blank Page ...
Страница 41: ...Revised 11 21 18 2 19 Overview Blank page ...
Страница 42: ...Nicolet Cortical Stimulator 2 20 Revised 11 21 18 ...
Страница 43: ...Revised 11 21 18 3 1 3 Performing a Cortical Stimulation Study ...
Страница 44: ...Nicolet Cortical Stimulator 3 2 Revised 11 21 18 Blank Page ...
Страница 69: ...Revised 11 21 18 4 1 4 ElectromagneticCompatibility EMC ...
Страница 70: ...Nicolet Cortical Stimulator 4 2 Revised 11 21 18 Blank Page ...
Страница 77: ...Revised 11 21 18 5 1 5 Error Codes ...
Страница 78: ...Nicolet Cortical Stimulator 5 2 Revised 11 21 18 Blank Page ...