Nicolet V Series Amplifiers
4-6
8/11/16
Calibration
The V Series amplifiers include a built-in calibration signal generator and the means to
connect this generator to the amplifier inputs at the first stage preamplifier. The generator
outputs a 5 Hz, 500
μ
V peak-to-peak square wave. Gain calibration and correction is
critical to the accurate creation of bipolar montages from the referentially recorded data.
IMPORTANT: It is recommended that a calibration procedure be performed prior
to every recording as standard practice.
NOTE:
Please refer to your host application for calibration instructions.
Event pushbutton
The optional Event pushbutton can be used to signal the recording software for event
marking.
The pushbutton and cabling can be unplugged and removed without affecting the normal
operation of the amplifier.
Photic stimulation
The V Series amplifiers include a built-in photic connection. Stimulation protocols can only be
setup and executed by the application software on the host computer. See the Host software
manual for details.
Содержание Nicolet V Series
Страница 2: ......
Страница 5: ...8 11 16 1 1 1 Introduction...
Страница 6: ...Nicolet V Series Amplifiers 1 2 August 11 2016 Blank page...
Страница 16: ...Nicolet V Series Amplifiers 1 12 August 11 2016 Blank page...
Страница 17: ...8 11 16 2 1 2 Product Description...
Страница 18: ...Nicolet V Series systems 2 2 8 11 16 Blank page...
Страница 26: ...Nicolet V Series systems 2 10 8 11 16 Blank page...
Страница 27: ...8 11 16 3 1 3 Installation...
Страница 28: ...Nicolet V Series systems 3 2 8 11 16 Blank page...
Страница 42: ...Nicolet V Series systems 3 16 8 11 16 Blank page...
Страница 43: ...8 11 16 4 1 4 Operation...
Страница 44: ...Nicolet V Series Amplifiers 4 2 8 11 16 Blank page...
Страница 49: ...8 11 16 5 1 5 Maintenance...
Страница 50: ...Nicolet VSeries Amplifiers 5 2 8 11 16 Blank page...
Страница 54: ...Nicolet V Series Amplifiers 2 8 11 16 Explosion hazard 3 4 Patient safety 3 5...