Nicolet V-Series systems
3-14
8/11/16
V-Series IP
address setup
NeuroWorks / SleepWorks software configuration
This section provides instructions on programming the v32 and v44 amplifiers to
communicate with the NeuroWorks/ SleepWorks software tool via either DHCP or TCP/
IP. If applicable, it is possible choose an IP address that is consistent with the existing
networking scheme of the hospital. The tool used for this is ProgramDevice.exe, which
allows you to configure the hardware protocol required for each amplifier and system.
1. Cycle amplifier power.
2. From the Neuroworks install directory folder, double-click on ProgramDevice.exe
(Neuroworks > ProgramDevice.exe).
DHCP programing
1. Check the box next to
Obtain an IP address automatically (DHCP)
.
This action will gray out the IP address, Subnet mask and Default gateway.
2. Click
Program Device
.
The "Device is configured successfully" notification window appears.
3. Click
Close
.
TCP/IP programing
1. Manually edit the IP address to the facilities desired address by clicking in the field
and typing the new
IP address
.
2. Click
Save
.
3. Click
Program Device
.
The "Device is configured successfully" notification window appears.
4. Click
Close
.
Finishing setup
1. On the desktop, double-click on the
Natus Database
shortcut.
2. Click
Tools > Options
.
3. Select the
Headboxes
tab.
4. Click
Refresh
.
5. All the amplifiers/headboxes connected to the network will be listed in the Headbox
Configuration area.
6. Click
OK
to exit the Natus Database Options window.
7. Click
Database > Exit
to close the main software application.
Содержание Nicolet V Series
Страница 2: ......
Страница 5: ...8 11 16 1 1 1 Introduction...
Страница 6: ...Nicolet V Series Amplifiers 1 2 August 11 2016 Blank page...
Страница 16: ...Nicolet V Series Amplifiers 1 12 August 11 2016 Blank page...
Страница 17: ...8 11 16 2 1 2 Product Description...
Страница 18: ...Nicolet V Series systems 2 2 8 11 16 Blank page...
Страница 26: ...Nicolet V Series systems 2 10 8 11 16 Blank page...
Страница 27: ...8 11 16 3 1 3 Installation...
Страница 28: ...Nicolet V Series systems 3 2 8 11 16 Blank page...
Страница 42: ...Nicolet V Series systems 3 16 8 11 16 Blank page...
Страница 43: ...8 11 16 4 1 4 Operation...
Страница 44: ...Nicolet V Series Amplifiers 4 2 8 11 16 Blank page...
Страница 49: ...8 11 16 5 1 5 Maintenance...
Страница 50: ...Nicolet VSeries Amplifiers 5 2 8 11 16 Blank page...
Страница 54: ...Nicolet V Series Amplifiers 2 8 11 16 Explosion hazard 3 4 Patient safety 3 5...