Function
Icon
Description
Curve Selection
(Curve
Selection)
Click to select the curve type:
•
THR
(THR) (Threshold level) (Tone)
•
MCL
(MCL) (Most Comfortable Loudness level)
•
UCL
(UCL) (Uncomfortable Loudness level)
•
SDT
(SDT) (Speech Detection Threshold) (Speech)
•
SRT
(SRT) (Speech Recognition Threshold) (Speech)
•
WRS/SRS
(WRS/SRS) (Word Recognition Score/Sentence Recognition
Score) (Speech)
Talk Forward
(Talk For-
ward)
The
Talk Forward
(Talk Forward) dialog is described in the MADSEN Aster-
a²Reference Manual.
Monitor/Level
(Mon-
itor/Level)
•
Monitor
(Monitor)
Click to enable monitoring of the stimulus channel.
•
Level
(Level)
Adjust the slider to set the preferred monitoring level for the respective
audiometer channels.
Test Selector
(Test
Selector)
The
Test Selector
(Test Selector) dialog is described in the OtosuiteUser
Guide.
11
Proper transducer placement
Headphones
1. Loosen the headband and place both the left and right side of the headphones simultaneously.
Note
•
If the headphones are not placed properly, there is risk of causing the ear canal to collapse which will result in
elevated thresholds.
2. Aim the center of the headphones towards the patient's ear canals and gently place them against the ears.
3. Tighten the headband while holding the headphones in place with your thumbs.
4. Examine the placement of the headphones to make sure they are level, and properly positioned.
Insert Earphones
Young children tolerate insert earphones better than headphones.
1. Select the largest foam eartip that will fit into the patient's ear.
If the eartip is too small the sound will leak out and the dB level will not be accurate at the eardrum.
Insert earphones have greater attenuation between ears especially at the low frequencies; this reduces the need for
masking.
MADSEN Astera²
17
11 Proper transducer placement