Hardware Instructions for Use
24
Patient Ground Connectors
Connect the patient's ground electrodes to the green
connectors.
WARNING
Do not connect the "patient ground" to the
protective earth connection on the Isolating Transformer
Unit, or to any other "ground" connections, as the electrode
inputs are galvanically isolated.
Patient Ground Connector
Connect the patient's ground electrode to the green
connector.
Rear Panel
LINK
HS Link Input
Connector
-
Main Unit - Side Panel
WARNING
Due to risk of electric shock, the operator
and / or patient must not directly or indirectly touch the
metal shield on the LINK cable attached to the rear of the
amplifier.