ALGO® 5 Newborn Hearing Screener Service Manual
P/N 000976 Rev.A
21
2.4.2.2.
ATA 5 Cable calibration due Warnings
The ALGO 5 checks the
Due for calibration
date of the ATA 5 each time the
screener is turned on. Different Warnings are issued related to the date the
screener is turned on. The procedure is described in FIGURE 14.
Time related Warnings in case ATA is due to calibration
Screening
Disabled
Due date -90 days
Warning 1
Warning 2
Due date
Due date + 90 days
Warning 1
: Warning:The Acoustic Transducer requires re-calibration on …
Warning 2:
Warning:Calibration has expired on .. . Please renew the calibration of the acoustic transducer!
Disabled:
Error: Calibration of the acoustic transducer has expired on .. . Please renew the calibration of the acoustic transducer!
Screening is no longer possible!
FIGURE 14: ATA 5 Cable calibration due Warnings
Annual calibration of the ATA 5 cable verifies and/or sets the intensity level of the click or
output stimulus.
ATA 5 cable calibration may only be performed by a qualified representative of Natus
Medical. For instructions on returning the ATA 5 cable to Natus Medical, contact Natus
Technical Service or your local distributor.