14
ALGO® 5 Newborn Hearing Screener Service Manual
P/N 000976 Rev.A
2.
Using an Ohm Meter, measure impedance between the positive and negative
contacts to get a baseline impedance.
3.
Using the Ohm Meter, measure the impedance from the ground pin on the outlet
plug to any screw on a cart column, such as the bracket that attaches the panel
PC to the column.
4.
Subtract the measured value from the baseline value measured in step 2 to get
the ground integrity value. Verify that the impedance is less than 100 milliohm
(0.1 ohm).
5.
Ground integrity should be checked whenever a shelf is adjusted in height.
100 milliohms or less
TABLE 1: Ground Integrity
2.3.2. Earth leakage current
AC
Polarity
Line cord
Neutral cord
IEC 60601-1
UL 60601-1
ANSI/NFPA 99
Normal
Closed
Closed
500 µA
300 µA
Reversed
Closed
Closed
500 µA
300 µA
Normal
Open
Closed
1000 µA
500 µA
Normal
Closed
Open
1000 µA
500 µA
TABLE 2: Earth leakage current
2.3.3. Enclosure leakage current
AC line
cord
polarity
AC line cord
neutral
conductor
AC line cord ground
conductor
IEC 60601-1
UL 60601-1
ANSI/NFPA 99
Normal
Closed
Closed
100 µA
300 µA
Normal
Closed
Open
500 µA
500 µA
Normal
Open
Closed
100 µA
300 µA
Reversed
Closed
Open
500 µA
500 µA
Reversed
Open
Closed
500 µA
300 µA
Reversed
Closed
Open
500 µA
500 µA
TABLE 3: Enclosure leakage current