ALGO® 5 Newborn Hearing Screener Service Manual
P/N 000976 Rev.A
13
CAUTION:
Press gently on the Panel PC display! Too much pressure can cause
permanent damage. Do not spray diluted cleaner directly on the display.
•
Clean the sensor and earphone cables with a cloth dampened with water,
isopropyl alcohol, or a 10% bleach solution.
Do not
stretch the cables.
Delicate shielding in the wires can be damaged, resulting in electrical shorting.
Refer to your hospital’s protocol for the frequency of cleaning equipment that
comes into contact with the patient.
•
Wipe off any remaining dirt or spots on the screener and cables with a cloth
dampened with a mild solution of commercial detergent, hospital cleaning
agent and water, or isopropyl alcohol.
CAUTION:
Do not spray or pour cleaners directly on the screener. Keep liquids,
including cleaning fluid, away from the screener’s keyboard, earphone
transducers, and other openings. Never spray cleaner directly onto the screener.
Never use harsh or abrasive chemical products to clean the screener. Do not
immerse patient cables or transducers in water or cleaning solutions.
2.2.2. Cleaning the cart
•
The cart may be cleaned with most mild, non-abrasive solutions commonly used
in the hospital environment (e.g., diluted bleach, ammonia, or alcohol solutions).
•
The surface finish will be permanently damaged by strong chemicals and solvents
such as acetone and trichloroethylene.
•
Abrasive material such as steel wool should never be used to clean any
component of the cart.
•
Never submerge the cart base or allow liquids to enter any component of the cart.
Wipe any cleaning agents off cart components immediately using a water-
dampened cloth. Dry all components thoroughly after cleaning.
•
Damage caused by the use of unapproved substances or processes will not be
covered by warranty. It is recommended that any cleaning solution be tested on
a small area of the cart that is not easily visible to verify compatibility.
CAUTION:
To clean or sterilize mounted instruments or accessory equipment,
refer to the specific instructions delivered with those products. Consult your
hospital’s infection control officer or epidemiologist.
2.3. Safety
testing
The safety tests meet the standards and are performed in accordance with, IEC
60601-1, (Amendment 1, Amendment 2,) and UL 60601, for instruments as Class
1 and Type BF and ANSI/NFPA Standard 99.
Applicable tests for these standards are listed in following sections (2.3.1-2.3.3).
Technicians must be familiar with the standards applicable to their institution and
country.
2.3.1. Ground
Integrity
1.
Unplug the ALGO 5 power cord from the wall outlet. Make sure it is a 3-prong
power cord (with ground).