Step 3.

   Select screening mode: 

Use the 

up/down arrow

 keys to select the screening mode. 

Below are the two choices (if SpeedScreen is enabled). Select the Enter Data and 
Screen Baby (ScreenBaby) mode. Press the 

OK

 key. 

a.  SpeedScreen allows you to immediately begin screening, with the option of 

entering infant data during the screening process. (This option is disabled at 
default.) 

b.  Enter Data/ScreenBaby allows you to create an infant record and select screening 

configurations prior to initiating the screening process.

 

Enter infant data in the Infant Data window using the 

toggle

 key. Use the 

up/down 

arrow

 keys to move from data field to data field (press and hold key for rapid 

scrolling).   

 

 

a.  There are three data fields in the first Infant Data Window. Infant Identifier, Last 

Name, and Frist Name. There are another three data fields in the second window. 
Date of Birth, User Identifier, and Comment.  

 
Press the 

OK

 key when data entry is complete. 

 

Step 4. 

Prepare baby and attach supplies:

 

Prepare the infant for screening by ensuring that 

no exclusion criteria are present, and that the infant is in a quiet state. Collect 
screening supplies: Natus single-use earphones (pair) and single-use sensor (triplet) 
for the infant.

 

a.  Attach sensor cable clips to sensors (clip onto the purple section of sensor).  See 

Figure 1 below.  

 

Figure 1 

 

Note: Clips may be attached after placement of sensors on infant.  

b.  Evaluate infant’s skin type at Vertex, Nape, and Common sites.   Clean or prep if 

necessary.  Use NuPrep gel or saline and gauze, if appropriate. 

Natus Medical • www.natus.com 

Page 6 of 12 

P/N 051179E 

Содержание ALGO 3i

Страница 1: ...h filter settings Note Ensure that the notch filter setting of 50Hz or 60Hz is correctly adjusted for your region An incorrect setting may result in high myogenic interference while screening 6 Set th...

Страница 2: ...OK Key Up Down Arrow Keys The display will always indicate to the user which keys are used to select scroll advance to the next window or go back to the previous window Examples Go Back to MainMenu Ad...

Страница 3: ...e at a time Multi choice options are indicated by a toggle key symbol See illustration below In this case the black select bar indicates use of the toggle keys to select or deselect the desired option...

Страница 4: ...e black select bar will indicate which data field is active and a cursor symbol will indicate the current position within the data field To begin scrolling through the keypad click on the up or down t...

Страница 5: ...up Warnings The ALGO 3i screener will briefly display a Start Up window while it performs a self check Warnings may appear which may require your acknowledgement The MainMenu will appear next Step 2...

Страница 6: ...There are three data fields in the first Infant Data Window Infant Identifier Last Name and Frist Name There are another three data fields in the second window Date of Birth User Identifier and Comme...

Страница 7: ...back Common site front or back of either shoulder not on cheek Note Ensure the tab portion of all sensors are pointed towards the same side of baby Note Hold sensors in place for a few seconds to war...

Страница 8: ...proceed 35 dBnHL screening only default setting 40 dBnHL screening option enabled Step 6 Start Screening Check Impedances Test immediately begins after screening parameters are selected Monitor the i...

Страница 9: ...reening in progress Monitor impedances and the Message window during the screening The ALGO 3i screener will alert you to any conditions which require your attention during the screening process The S...

Страница 10: ...nnected Battery charger connected Screening time limit exceeded Battery too low to complete screening Step 8 Observe screening complete At the completion of the screen the ALGO 3i screener will save a...

Страница 11: ...reening in Progress window i If impedances are between 8 kOhms and 11 kOhms the Impedances Marginal message will appear showing actual impedances and asking you to confirm your intention to begin scre...

Страница 12: ...his area Ensure that all infant data is accurate before clicking the OK key e If you do not wish to enter infant data for the screening record proceed with screening At the completion of the screening...

Отзывы: