natus ALGO 3i Скачать руководство пользователя страница 12

Natus Medical • www.natus.com 

Page 12 of 12 

P/N 051179E 

Note:

 Once you exit the Infant Data window, you may not re-enter this area. 

Ensure that all infant data is accurate before clicking the 

OK

 key. 

e.  If you do not wish to enter infant data for the screening record, proceed with 

screening. 
At the completion of the screening, the device will save the screening record and will 
display the screening results. The infant data fields will be blank. 

 

 

 

 

Содержание ALGO 3i

Страница 1: ...h filter settings Note Ensure that the notch filter setting of 50Hz or 60Hz is correctly adjusted for your region An incorrect setting may result in high myogenic interference while screening 6 Set th...

Страница 2: ...OK Key Up Down Arrow Keys The display will always indicate to the user which keys are used to select scroll advance to the next window or go back to the previous window Examples Go Back to MainMenu Ad...

Страница 3: ...e at a time Multi choice options are indicated by a toggle key symbol See illustration below In this case the black select bar indicates use of the toggle keys to select or deselect the desired option...

Страница 4: ...e black select bar will indicate which data field is active and a cursor symbol will indicate the current position within the data field To begin scrolling through the keypad click on the up or down t...

Страница 5: ...up Warnings The ALGO 3i screener will briefly display a Start Up window while it performs a self check Warnings may appear which may require your acknowledgement The MainMenu will appear next Step 2...

Страница 6: ...There are three data fields in the first Infant Data Window Infant Identifier Last Name and Frist Name There are another three data fields in the second window Date of Birth User Identifier and Comme...

Страница 7: ...back Common site front or back of either shoulder not on cheek Note Ensure the tab portion of all sensors are pointed towards the same side of baby Note Hold sensors in place for a few seconds to war...

Страница 8: ...proceed 35 dBnHL screening only default setting 40 dBnHL screening option enabled Step 6 Start Screening Check Impedances Test immediately begins after screening parameters are selected Monitor the i...

Страница 9: ...reening in progress Monitor impedances and the Message window during the screening The ALGO 3i screener will alert you to any conditions which require your attention during the screening process The S...

Страница 10: ...nnected Battery charger connected Screening time limit exceeded Battery too low to complete screening Step 8 Observe screening complete At the completion of the screen the ALGO 3i screener will save a...

Страница 11: ...reening in Progress window i If impedances are between 8 kOhms and 11 kOhms the Impedances Marginal message will appear showing actual impedances and asking you to confirm your intention to begin scre...

Страница 12: ...his area Ensure that all infant data is accurate before clicking the OK key e If you do not wish to enter infant data for the screening record proceed with screening At the completion of the screening...