Quick Reference Guide for using the  

ALGO

®

 3i Newborn Hearing Screener 

 

 

A.  Power-On slider switch 
B. Display 
C.  Navigation buttons (keys) 
D.  Earphone (ATA) cable connection  
E.  Sensor (Preamp/PCA) cable 

connection 

F.  Infrared (IR) port: printing/data 

export 

G.  DC power port: battery charger 

connection 

H.  SoftClip™: device stand/bassinet 

clip

 

 

Initial Set Up Of The Screener 

Note:

 Refer to the ALGO 3i user manual, Chapter VIII: Service Manual: Description and Initial 

Set-up of Device Components.

 

1.  Unpack the ALGO 3i screening system and its accessories. 
2.  Set up the screener properly, attaching the earphone and sensor cables (if unattached). 
3.  Power on the ALGO 3i screener and fully charge it for 6 hours. 
4.  After fully charging the battery, ensure that the correct time and date are set. 
5.  Change the User Preferences and/or Biomed Preferences settings to meet your program 

requirements, i.e. time/date format, language, alpha-numeric entry, or notch filter 
settings.  

Note:

 Ensure that the notch filter setting of 50Hz or 60Hz is correctly adjusted for your 

region.  An incorrect setting may result in high myogenic interference while screening.

 

6.  Set the ALGO 3i system in an appropriate location for screening. 

 

Selecting A Baby 

 

Indications 

for 

use:          

• 

Term, healthy newborns 

• 

NICU graduates, 34 weeks gestational 
age and under 6 months of age 

• 

Newborns with normal outer ear 
anatomy and no obvious deformities of 
the head or neck 

• 

Ready for discharge, or scheduled to go 
home within a few days 

• 

Sleeping or in a relaxed state (usually 
after a recent feeding) 

• 

In an open crib 

 
 

Exclusion criteria:  

 

 

 

 

 

 

 

 

 

• 

Infants not between 34 weeks 
gestational and 6 months of age 

• 

Infants on ventilators or in incubators 

• 

Infants on CNS (central nervous 
system) stimulants 

• 

Infants receiving ototoxic medications 

• 

Infants with compromised skin or 
jaundice 

• 

Infants exhibiting cranio-facial head 
abnormalities, including structural 
abnormalities of the outer, middle, or 
inner ear 

• 

Infants with known or suspected 
neurologic conditions 

• 

Infants in an agitated state 

Natus Medical • www.natus.com 

Page 1 of 12 

P/N 051179E

 

Содержание ALGO 3i

Страница 1: ...h filter settings Note Ensure that the notch filter setting of 50Hz or 60Hz is correctly adjusted for your region An incorrect setting may result in high myogenic interference while screening 6 Set th...

Страница 2: ...OK Key Up Down Arrow Keys The display will always indicate to the user which keys are used to select scroll advance to the next window or go back to the previous window Examples Go Back to MainMenu Ad...

Страница 3: ...e at a time Multi choice options are indicated by a toggle key symbol See illustration below In this case the black select bar indicates use of the toggle keys to select or deselect the desired option...

Страница 4: ...e black select bar will indicate which data field is active and a cursor symbol will indicate the current position within the data field To begin scrolling through the keypad click on the up or down t...

Страница 5: ...up Warnings The ALGO 3i screener will briefly display a Start Up window while it performs a self check Warnings may appear which may require your acknowledgement The MainMenu will appear next Step 2...

Страница 6: ...There are three data fields in the first Infant Data Window Infant Identifier Last Name and Frist Name There are another three data fields in the second window Date of Birth User Identifier and Comme...

Страница 7: ...back Common site front or back of either shoulder not on cheek Note Ensure the tab portion of all sensors are pointed towards the same side of baby Note Hold sensors in place for a few seconds to war...

Страница 8: ...proceed 35 dBnHL screening only default setting 40 dBnHL screening option enabled Step 6 Start Screening Check Impedances Test immediately begins after screening parameters are selected Monitor the i...

Страница 9: ...reening in progress Monitor impedances and the Message window during the screening The ALGO 3i screener will alert you to any conditions which require your attention during the screening process The S...

Страница 10: ...nnected Battery charger connected Screening time limit exceeded Battery too low to complete screening Step 8 Observe screening complete At the completion of the screen the ALGO 3i screener will save a...

Страница 11: ...reening in Progress window i If impedances are between 8 kOhms and 11 kOhms the Impedances Marginal message will appear showing actual impedances and asking you to confirm your intention to begin scre...

Страница 12: ...his area Ensure that all infant data is accurate before clicking the OK key e If you do not wish to enter infant data for the screening record proceed with screening At the completion of the screening...

Отзывы: