IX. Reference
ALGO
®
3i Newborn Hearing Screener IFU 024913 RevD Revision Date 08/05/2020
49
Table 9 - Test specifications for ENCLOSURE PORT IMMUNITY to RF wireless
communications equipment
Test
Frequency
(MHz)
Band
(MHz)
Service
Modulation
Maximum
Power
(W)
Distance
(m)
Immunity
Test Level
(V/m)
385
380 – 390
TETRA 400
Pulse modulation 18 Hz
1,8
0,3
27
450
430 – 470
GMRS 460,
FRS 460
FM
± 5 kHz deviation
1 kHz sine
2
0,3
28
710
704 – 787
LTE Band 13, 17
Pulse modulation 217 Hz
0,2
0,3
9
745
780
810
800 – 960
GSM 800/900
TETRA 800
iDEN 820
CDMA 850
LTE Band 5
Pulse modulation 18 Hz
2
0,3
28
870
930
1,720
1,700 - 1,990
GSM 1800
CDMA 1900
GSM 1900
DECT
LTE Band 1, 3, 4,
25; UMTS
Pulse modulation 217 Hz
2
0,3
28
1,845
1,970
2,450
2,400 – 2,570
990Bluetooth
WLAN
802.11 b/g/n
RFID 2450
LTE Band 7
Pulse modulation 217 Hz
2
0,3
28
5,240
5,100 – 5,800
WLAN 802.11 a/n
Pulse modulation 217 Hz
0,2
0,3
9
5,500
5,785