If you are using high voltage (

30 V

rms

 and 42.4 V

peak

 or 60 VDC), please

read the following warnings:

Warning:

If high voltages (

≥ 

≥ 

≥ 

≥ 

30 V

rms

 and 42.4 V

peak

 or 60 VDC) are present, YOU

MUST CONNECT THE SAFETY EARTH GROUND TO THE
GROUNDING LUG. This complies with safety agency requirements
from UL 1244 and IEC 1010 and protects against electric shock when
the terminal block is not connected to the VXIbus chassis. To connect
the safety earth ground to the grounding lug, run an earth ground wire
in the cable from the signal source to the terminal block. National
Instruments is NOT liable for any damages or injuries resulting from
inadequate safety earth ground connections.

Equipment described in this guide must be used in an Installation
Category II environment per IEC 664. This category requires local level
supply mains-connected installation. Do not use this equipment to
measure voltages higher than 300 V.

SHOCK HAZARD: Only qualified personnel aware of the dangers
involved should open this unit. Disconnect all power before removing
the cover. If signal wires are connected to the module or terminal block,
dangerous voltages may exist even when the equipment is turned off.
Before you remove any installed terminal block, disconnect the AC
power line or any high voltage sources (

≥≥≥≥

30 V

rms

 and 42.4 V

peak

 or

60 VDC) that may be connected to the terminal block.

When using the terminal block with high common-mode voltages, you
MUST insulate your signal wires to prevent electric shock by using
300 V wire minimum. National Instruments is NOT liable for any
damages or injuries resulting from inadequate signal wire insulation.

The chassis GND terminals on the terminal block are for grounding
high-impedance sources such as floating source (1 mA maximum). Do
NOT use these terminals as safety earth grounds.

Содержание VXI-TB-1000

Страница 1: ...ds You can configure these terminal boards to match any VXI DAQ module configuration The VXI TB 1000 populated with terminal boards can easily accommodate thermocouples RTDs strain gauges thermistors millivolt sources volt sources and current loop receivers What You Need to Get Started You need the following items to set up and use your VXI TB 1000 terminal board carrier One or more VXI TB 1000 te...

Страница 2: ...l level supply mains connected installation Do not use this equipment to measure voltages higher than 300 V SHOCK HAZARD Only qualified personnel aware of the dangers involved should open this unit Disconnect all power before removing the cover If signal wires are connected to the module or terminal block dangerous voltages may exist even when the equipment is turned off Before you remove any inst...

Страница 3: ... terminal board carrier 4 Install and tighten the four board screws 8 included with the board using a Phillips head number 1 screwdriver 5 If you are using only one terminal board you need to install the safety shield 10 included with the VXI TB 1000 Place the safety shield over the screw bosses 15 and tighten the two safety shield screws 4 included with the shield to the screw bosses Warning You ...

Страница 4: ... 3 1 1 Handle Recess Area 5 Cover 9 Screw Terminal 13 Cover Insulator 2 Mounting Screw 6 Cover Screw 10 Safety Shield 14 User Help Card 3 Grounding Lug 7 Terminal Board 11 Strain Relief Bar Screw 15 Screw Bosses 4 Safety Shield Screw 8 Board Screw 12 Strain Relief Bar Figure 1 VXI TB 1000 Parts Locator Diagram ...

Страница 5: ...move the strain relief bar 12 for easier access 3 Use a wire cutter and wire insulation stripper to strip the wire ends as necessary to connect them to screw terminals 4 Loosen the screws in the screw terminals 9 with a 0 09 in flathead screwdriver 5 Insert the stripped wires into the screw terminals Tighten the screws with a 0 09 in flathead screwdriver 6 Replace the strain relief bar if necessar...

Страница 6: ...o items in Figure 3 Note Before attaching the terminal block onto the VXI DAQ module front panel connector make sure the VXI DAQ module attachment screws 1 are tightened to the VXIbus chassis 1 Guide the terminal block onto the VXI DAQ module front panel connector 7 by aligning the terminal block handle recess area 3 to the ejector handles 2 on the VXI DAQ module 2 Connect the VXI DAQ module front...

Страница 7: ...4 8 2 1 Attachment Screw 3 Handle Recess Area 5 Mounting Screw 7 Front Panel Connector 2 Ejector Handle 4 VXI TB 1000 6 Mating Connector 8 VXI DAQ Module Figure 3 Installing the Terminal Block on the VXI DAQ Module ...

Отзывы: