Appendix A Specifications
© National Instruments Corporation
A-5
VXI-1500 User Manual
Safety
Table A-4.
Safety Specifications
Characteristic
Description
Safety Characteristics
EN61010-1 (UL 3111-1, c-UL)
IEC Characteristics
Installation Category II
Pollution Degree 2
Safety Class 1
Environmental
Table A-5.
Environmental Specifications
Characteristic
Description
Operating Temperature
0
o
to 40
o
C at full power
0
o
to 55
o
C at half power
Storage Temperature
-20
o
to 65
o
C
Operating Relative Humidity
10% to 95% RH non-condensing
Altitude
1.24 miles (2 km)
Functional Shock (Operating)
MIL-T-28800E CLASS 3
(Half-Sine Shock Pulse, 11 ms
duration, 30 g peak)
Random Vibration (Operational)*
5 to 500 Hz, 0.31 g
Random Vibration
(Non-Operational)*
5 to 500 Hz, 2.5 g
Electromagnetic Compatibility
(EMC) Emissions
Refer to DOC supplied with
chassis for compliance to
relevant directives.
EMI
FCC Class A compliant
*
Random vibration profiles were developed in accordance with
MIL-STD-810E Method 514. Test levels exceed those
recommended in MIL-STD-810E for Category 1 (Basic
Transportation, Figures 514.4-1 through 514.4-3.