36
|
ni.com
|
NI VirtualBench Calibration Procedure
22.
(VB-8034/8054)
Remove the connections from channel 3. Connect VirtualBench channel
4 to the oscilloscope calibrator active head, as shown in Figure 18.
Figure 18.
VB-8034/8054 Oscilloscope Gain Channel 4 Connection Diagram
23.
(VB-8034/8054)
Repeat steps 2 to 16 for all ranges on channel 4.
Note
(VB-8034/8054)
When calling
MSO Configure Analog Channel
in
step 4, change the
Channel
input to
mso/4
.
24.
(VB-8034/8054)
Remove the connections from channel 4.
1
VirtualBench Oscilloscope
2
Oscilloscope Calibrator
3
Oscilloscope Calibrator Active Head
2
CH 2
CH 1
1
3