 National Instruments Corporation

15

NI USB-9237 User Guide and Specifications

Each channel on the NI USB-9237 has an independent 24-bit ADC and 
input amplifier that enable you to sample signals from all four channels 
simultaneously. 

The NI USB-9237 also includes filters to prevent aliasing. The filters on the 
NI USB-9237 filter according to the sampling rate.

Understanding NI USB-9237 Filtering

The NI USB-9237 uses a combination of analog and digital filtering to 
provide an accurate representation of desirable signals while rejecting 
out-of-band signals. The filters discriminate between signals based on the 
frequency range, or bandwidth, of the signal. The three important 
bandwidths to consider are the passband, the stopband, and the alias-free 
bandwidth. 

The NI USB-9237 represents signals within the passband as accurately 
as possible, as quantified primarily by passband flatness and phase 
nonlinearity. The filters reject frequencies within the stopband as much as 
possible, as quantified by stopband rejection. All signals that appear in the 
alias-free bandwidth are either unaliased signals or signals that have been 
filtered by at least the amount of the stopband rejection.

Passband

The signals within the passband have frequency-dependent gain or 
attenuation. The small amount of variation in gain with frequency is called 
the passband flatness. The filters of the NI USB-9237 adjust the frequency 
range of the passband to match the data rate. Therefore, the amount of gain 
or attenuation at a given frequency depends on the data rate. Figure 9 shows 
typical passband flatness for a range of data rates.

Figure 9.  

Typical Passband Response

 

Frequency/Sample Rate (kHz)

1

0.8

0.025

0.000

–0.025

–0.050

0.6

0.4

0.2

0

Gain (dB)

Содержание USB-9237

Страница 1: ...ations The NI USB 9237 provides a USB interface for four channels of 24 bit half full bridge analog input Introduction The NI USB 9237 consists of two components an NI 9237 module and an NI USB 9162 c...

Страница 2: ...mited Refer to the Specifications section for more details Hot Surface This icon denotes that the component may be hot Touching this component may result in bodily injury Caution Do not disconnect I O...

Страница 3: ...ming environments Select Start All Programs National Instruments NI DAQ NI DAQmx Help LabVIEW If you are a new user use the Getting Started with LabVIEW manual to familiarize yourself with the LabVIEW...

Страница 4: ...NET you can interactively create channels and tasks by launching the DAQ Assistant from MAX or from within Visual Studio NET You can generate the configuration code based on your task or channel in Me...

Страница 5: ...dio NET documentation To view the NI DAQmx NET documentation go to Start Programs National Instruments NI DAQ NI DAQmx NET Reference Help Expand NI Measurement Studio Help NI Measurement Studio NET Cl...

Страница 6: ...stallation instructions that accompany your software Example Programs The NI DAQmx CD contains example programs that you can use to get started programming with the NI USB 9237 Refer to the NI DAQmx f...

Страница 7: ...th the carrier as shown in Figure 3 Figure 3 Module Installation 4 Squeeze the latches and insert the NI 9237 module into the USB 9162 carrier 5 Press firmly on the connector side of the NI 9237 modul...

Страница 8: ...ormation LED Indicator The NI USB 9237 device has a green LED next to the USB connector The LED indicator indicates device status as listed in Table 1 When the device is connected to a USB port the LE...

Страница 9: ...can use to connect an external excitation voltage source to the module Table 2 lists the connections between an external excitation voltage source and the NI USB 9237 Note For a list of accessories a...

Страница 10: ...e 2 Terminal Assignments RJ 50 10p10c Modular Plug and Receptacle Pin Numbers RJ 50 Pin RJ 45 Pin Signal Name Signal Description 1 SC Shunt calibration 2 1 AI Positive input signal 3 2 AI Negative inp...

Страница 11: ...quarter bridge with the NI USB 9237 if you either add a resistor externally to create a half bridge or use the NI 9944 or NI 9945 Quarter Bridge Completion Accessory Visit ni com and search for the NI...

Страница 12: ...alibration Remote Sensing Remote sensing continuously and automatically corrects for errors in excitation leads and generally is most appropriate for full and half bridge sensors Long wire and smaller...

Страница 13: ...ls connected to the sensor and then apply or remove the shunt calibration resistance in software While remote sensing corrects for resistances from the EX terminals on the NI USB 9237 to the sensor sh...

Страница 14: ...ment The 150 mW limit allows you to power full and half bridges as follows Four 350 half bridges at 5 0 V Four 350 full bridges at 3 3 V Four 120 half bridges at 2 5 V If you need an excitation voltag...

Страница 15: ...he NI USB 9237 represents signals within the passband as accurately as possible as quantified primarily by passband flatness and phase nonlinearity The filters reject frequencies within the stopband a...

Страница 16: ...9237 is not an aliased artifact of signals at a higher frequency The alias free bandwidth is defined by the ability of the filter to reject frequencies above the stopband frequency and equals the dat...

Страница 17: ...brated Phase nonlinerarity 0 to 1 kHz 0 001 0 to 20 kHz 0 1 Input delay 4 8 s 38 4 fs Passband Frequency 0 45 fs Flatness 0 1 dB max Error Percent of Reading Gain Error Percent of Range Offset Error C...

Страница 18: ...rth ground2 140 dB 0 to 60 Hz Relative to EX 85 dB 0 to 1 kH Input noise Spurious free dynamic range SFDR 106 dB 1 KHz 60 dBFS Total harmonic distortion fundamental 20 dBFS 1 kHz 100 dB 8 kHz 90 dB Ex...

Страница 19: ...power 150 mW max External voltage 2 V to 10 V Power Requirements Current consumption from USB 500 mA max Suspend mode 2 5 mA max Bus Interface USB specification USB 2 0 Hi Speed Physical Characterist...

Страница 20: ...by regulated low voltage sources and electronics Do not connect the NI USB 9237 to signals or use for measurements within Measurement Categories II III or IV Safety Standards The NI USB 9237 is desig...

Страница 21: ...munity EN 55011 Emissions Group 1 Class A CE C Tick ICES and FCC Part 15 Emissions Class A Note For EMC compliance operate this device with double shielded cables CE Compliance This product meets the...

Страница 22: ...egulations and directives with which NI complies as well as other environmental information not included in this document Waste Electrical and Electronic Equipment WEEE EU Customers At the end of thei...

Страница 23: ...onal Instruments also has offices located around the world to help address your support needs For telephone support in the United States create your service request at ni com support and follow the ca...

Отзывы: