Appendix A
Specifications
©
National Instruments Corporation
A-3
Random vibration
Operating ........................................ 5 to 500 Hz, 0.3 g
rms
(with solid-state hard drive)
Nonoperating .................................. 5 to 500 Hz, 2.4 g
rms
(Tested in accordance with IEC-60068-2-64. Nonoperating test
profile exceeds the requirements of MIL-PRF-28800F, Class B.)
Note
Specifications are subject to change without notice.
Safety
This product is designed to meet the requirements of the following
standards of safety for electrical equipment for measurement, control,
and laboratory use:
•
IEC 61010-1, EN 61010-1
•
UL 61010-1, CSA 61010-1
Note
For UL and other safety certifications, refer to the product label or the
Electromagnetic Compatibility
This product is designed to meet the requirements of the following
standards of EMC for electrical equipment for measurement, control,
and laboratory use:
•
EN 61326 (IEC 61326): Class A emissions; Basic immunity
•
EN 55011 (CISPR 11): Group 1, Class A emissions
•
AS/NZS CISPR 11: Group 1, Class A emissions
•
FCC 47 CFR Part 15B: Class A emissions
•
ICES-001: Class A emissions
Note
For the standards applied to assess the EMC of this product, refer to the
section.
Note
For EMC compliance, operate this device with shielded cabling.