TBX-1303 32-Channel Isothermal Terminal Block
4
ni.com
Installing the TBX-1303
Perform the following steps to mount the SH96-96 cable assembly and
connect the TBX-1303 to the SCXI module. Refer to Figures 1 and 2
as needed.
Caution
Refer to the
Connecting Signals
section before connecting the signals. If signal
wires are connected to the terminal block, dangerous voltages can exist even when the
equipment is powered off.
1.
Power off the SCXI chassis.
2.
Power off the computer that contains the E Series data acquisition
(DAQ) device, or disconnect the device from the SCXI chassis.
3.
Connect the TBX cable adapter to the SCXI module, and secure
the adapter by tightening both thumbscrews.
Figure 1.
Connecting the SH96-96 Cable to the SCXI Module
1
Backshell Mounting Screws
2
SH96-96 Cable
3
Thumbscrews
4
SCXI Chassis
5
TBX Cable Adapter
3
4
3
1
5
1
2