Appendix C
Common Questions
SCXI-1581 User Manual
C-2
ni.com
Can the SCXI-1581 current outputs be interactively controlled in
MAX or programmatically controlled using NI-DAQ function calls,
LabVIEW, or Measurement Studio?
No. The current-output level is 100
µ
A as long as the chassis is powered on.
You cannot power off or adjust the current output using MAX, NI-DAQ
function calls, or an ADE such as LabVIEW or Measurement Studio. If you
require this functionality, consider using an SCXI-1124 module or NI 670
X
device instead.
Can I create virtual channels for an SCXI-1581 current-output
channel?
No. As discussed above, you cannot programmatically control the current
outputs.
Can I connect
N
current-output channels in parallel to create a
precision current source that provides
N
×
100
µ
A?
This is not recommended. The accuracy or stability of the resulting
current-output value is unpredictable.
Can I connect
N
current-output channels in series to achieve a higher
terminal-voltage compliance limit?
No. Each current source is ground referenced. Therefore, you cannot place
multiple current-outputs in series.
Are the SCXI-1581 current output channels isolated with respect to
each other, the E Series DAQ device, or ground?
No. The SCXI-1581 does not contain any isolation circuitry. All current
outputs are referenced to chassis ground. If you require isolated current
outputs, consider the SCXI-1124 module instead.
Can I modify the SCXI-1581 circuitry to generate current at a level
different than 100
µ
A?
No. Do not attempt to modify any circuitry in the SCXI-1581.
Are there any user-serviceable parts inside the SCXI-1581?
No. There are no fuses, potentiometers, switches, socketed resistors, or
jumpers inside the module. Disassembly of the module for any reason can
void its warranty and nullify its accuracy specification.