Specifications
All specifications are typical at 25 °C unless otherwise specified.
Maximum Voltage
Terminal to earth ....................................150 V
rms
or VDC
Terminal to terminal ...............................150 V
rms
or VDC
Environment
Operating temperature ............................0 to 50 °C
Storage temperature ................................–20 to 70 °C
Relative humidity ...................................10 to 90%
Safety
Designed in accordance with IEC61010-1, UL 3111-1, and
CAN/CSA C22.2 No. 1010.1 for electrical measuring and test equipment
For use at altitudes up to 2000 m
Indoor use only
Installation Category I
Pollution Degree 2
Emissions and Immunity
EMC/EMI ...............................................CE, C-Tick and FCC Part 15
(Class A) Complaint
Electrical emissions ................................EN 55011 Class A at 10 m;
FCC Part 15A above 1 GHz
Electrical immunity ................................Evaluated to EN 61326:1998,
Table 1
Note
For full EMC and EMI compliance, you must operate this device with shielded
cabling. See the Declaration of Conformity (DoC) for this product for any additional
regulatory compliance information. To obtain the DoC for this product, click
Declaration of Conformity
at
ni.com/hardref.nsf
. This website lists the DoCs by
product family. Select the appropriate product family, followed by your product, and a link