SCXI Quick Start Guide

18

ni.com

Step 14. Take an NI-DAQmx Measurement

This step applies only if you are programming your device using NI-DAQ or NI application 
software. Refer to 

Take an NI-DAQmx Measurement 

in the 

DAQ Getting Started Guide 

for 

information.

Use Your Task in an Application

Refer to the 

DAQ Getting Started Guide

 for information.

Troubleshooting

This section contains troubleshooting tips and answers to questions that SCXI users commonly 
ask NI technical support staff. 

Tips

Before you contact NI, try the following troubleshooting tips:

If you have problems installing your software, go to 

ni.com/support/daqmx

. For 

hardware troubleshooting, go to 

ni.com/support

, enter your device name, or go to 

ni.com/kb

.

Go to 

ni.com/info

 and enter 

rddq8x

 for a complete listing of NI-DAQmx documents 

and their locations.

If you need to return your National Instruments hardware for repair or device calibration, 
refer to 

ni.com/info

 and enter the info code 

rdsenn

 to start the Return Merchandise 

Authorization (RMA) process.

Make sure the SCXI chassis is powered on. If you are using a PXI/SCXI combination 
chassis, make sure the PXI chassis is powered on.

Make sure you have installed the latest version of the NI-DAQ driver software that supports 
the devices in your system.

If MAX cannot establish communication with the chassis, try one or all of the following:

Connect the DAQ device to a different module in the chassis.

Try a different cable assembly.

Try a different chassis.

Try a different DAQ device.

Make sure that each SCXI chassis connected to a single DAQ device has a unique address.

Make sure that the cable is securely connected to the chassis.

Check for bent pins on the module, chassis backplane, and the device connector.

If you have multiple SCXI modules, remove all the modules and test each module 
individually.

If you are getting erroneous readings from the signal source, disconnect the signal source 
and short-circuit the input channel to ground. You should get a 0 V reading. Alternatively, 
connect a battery or other known signal source to the input channel.

Run an example program to see if you still get erroneous results.

Содержание SCXI-116 Series

Страница 1: ...SCXI 1143...

Страница 2: ...ating properly and set up multichassis systems It also describes the NI DAQmx software relative to SCXI and integrated signal conditioning products This document assumes you have already installed con...

Страница 3: ...liers Figure 1 SCXI System Components This icon denotes a caution which advises you of precautions to take to avoid injury data loss or a system crash When this symbol is marked on a product refer to...

Страница 4: ...section after installing a DAQ device or an SCXI USB device SCXI Chassis 1 Power off and unplug the chassis 2 Set the chassis address if your chassis is addressable Some older chassis are not address...

Страница 5: ...shown in the off position 3 Set the voltage selection tumbler to the correct voltage for your application Refer to the chassis documentation for more information 4 Connect the power cord Figure 4 PXI...

Страница 6: ...the PXI chassis complete the following steps 1 Touch any metal part of the chassis to discharge static electricity 2 Place the module edges into the top and bottom PXI module guides as shown in Figure...

Страница 7: ...n SCXI system and a USB compatible computer or USB hub so no intermediate DAQ device is needed SCXI USB modules such as the SCXI 1600 cannot be used in a PXI SCXI combination chassis or in multichassi...

Страница 8: ...pinout locations For information about sensors refer to ni com sensors For information about IEEE 1451 4 TEDS smart sensors refer to ni com teds 1 New SCXI Module 2 Existing SCXI Module 3 SCXI Chassi...

Страница 9: ...adapter such as the SCXI 1349 If there is an analog input module with simultaneous sampling capability in the chassis you must connect that module to the cable assembly or an error message appears ea...

Страница 10: ...annot have an external cable inserted into its rear 50 pin connector SCXI 1000 chassis through revision D do not have address jumpers or switches and respond to any address but you cannot use them in...

Страница 11: ...5 Install the SCXI 1349 cable adapter into the last SCXI chassis in the chain Refer to step 1 of the previous Single Chassis System section for instructions on installing the SCXI 1349 Step 8 Connect...

Страница 12: ...I System Step 9 Power On the SCXI Chassis If you are using an SCXI chassis the chassis power switch is shown in Figure 3 If you are using a PXI SCXI combination chassis the PXI and chassis power switc...

Страница 13: ...ches LabVIEW Choose this option if you have already configured your device in MAX Run Test Panels Launches MAX test panels for your device Configure and Test This Device Opens MAX Take No Action Recog...

Страница 14: ...Chassis If you installed an SCXI USB module such as the SCXI 1600 skip to Step 12 Configure the Chassis and Modules The SCXI USB module and associated chassis appear automatically under Devices and In...

Страница 15: ...he step numbers 1 Select the DAQ device cabled to the communicating SCXI module from Chassis Communicator If MAX detects only one DAQ device the device is chosen by default and this option is disabled...

Страница 16: ...put module in parallel mode in a multichassis configuration or another special configuration click the Cabling tab to adjust the settings for cabling If you are using standard multiplexed mode operati...

Страница 17: ...assis A message at the top of the SCXI Chassis Configuration window displays the status of the configuration You cannot save the chassis configuration if an error appears until you finish entering mod...

Страница 18: ...for NI DAQmx by selecting Help Help Topics NI DAQ MAX Help for NI DAQmx for instructions about creating NI DAQmx simulated devices and importing NI DAQmx simulated device configurations to physical de...

Страница 19: ...tart the Return Merchandise Authorization RMA process Make sure the SCXI chassis is powered on If you are using a PXI SCXI combination chassis make sure the PXI chassis is powered on Make sure you hav...

Страница 20: ...is for the positive analog supply and the fuse marked with a copper is for the negative analog supply 4 Using long nose pliers carefully remove the fuse 5 Take a new fuse and bend its leads so the com...

Страница 21: ...ith a thermocouple How do I stop the thermocouple reading from fluctuating Average the temperature readings to minimize fluctuations Also ensure proper field wiring techniques Most thermocouples are f...

Страница 22: ...1 Class A emissions FCC 47 CFR Part 15B Class A emissions ICES 001 Class A emissions Note For the standards applied to assess the EMC of this product refer to the Online Product Certification section...

Страница 23: ...nvironmental regulations and directives with which NI complies as well as other environmental information not included in this document Waste Electrical and Electronic Equipment WEEE EU Customers At t...

Страница 24: ...red trademark of Microsoft Corporation in the United States and other countries Other product and company names mentioned herein are trademarks or trade names of their respective companies For patents...

Отзывы: