Appendix A
Specifications
©
National Instruments Corporation
A-7
Maximum Working Voltage
Maximum working voltage refers to the signal voltage plus the
common-mode voltage.
Channel-to-earth..................................... 250 V, Installation Category II
Channel-to-channel ................................ 250 V, Installation Category II
Environmental
Operating temperature............................ 0 to 50 °C
Storage temperature ............................... –20 to 70 °C
Humidity ................................................ 10 to 90% RH, noncondensing
Maximum altitude .................................. 2,000 m
Pollution Degree (indoor use only) ........ 2
Safety
This product is designed to meet the requirements of the following
standards of safety for electrical equipment for measurement, control, and
laboratory use:
•
EC 61010-1, EN 61010-1
•
UL 3111-1, UL 61010B-1
•
CAN/CSA C22.2 No. 1010.1
Note
For UL and other safety certifications, refer to the product label or to
ni.com
.
Electromagnetic Compatibility
Emissions ............................................... EN 55011 Class A at 10 m. FCC
Part 15A above 1 GHz
Immunity ................................................ EN 61326:1997 + A2:2001,
Table 1
EMC/EMI............................................... CE, C-Tick and FCC Part 15
(Class A) Compliant
Note
For EMC compliance, operate this device with shielded cabling.