Chapter 3
Configuring Your SCXI Hardware and Software
©
National Instruments Corporation
3-13
Your SCXI-1120D is shipped in the 4.5 kHz position. Verify that both
stages are set to the same bandwidth to ensure that the required
bandwidth is achieved. Notice that one jumper block is available for
each filter stage.
3.
Configure the temperature sensor signal routing, if necessary.
Jumper W41 controls the routing of the temperature sensor signal
available on the SCXI-1320 terminal block. Use position 1-2
(the factory-default position) for MTEMP (multiplexed) mode.
Position 2-3 for DTEMP (direct) mode is no longer supported by the
current version of the SCXI-1320 terminal block.
4.
Configure the amplifier output referencing mode.
The signal output referencing is relevant only if you are connecting
this SCXI-1120/D module directly to a DAQ device.
Use jumper W46 to configure the various referencing modes of the
SCXI-1120/D instrumentation amplifier output. The correct output
referencing mode depends on the input mode of the DAQ device
you use.
3
W20-A
W20-B
W31
W32
4
W21-A
W21-B
W33
W34
5
W22-A
W22-B
W35
W36
6
W23-A
W23-B
W37
W38
7
W24-A
W24-B
W39
W40
Table 3-7.
SCXI-1120D Filter Jumper Allocation (Continued)
Input
Channel
Number
First
Filter Jumper
Second
Filter Jumper
4.5 kHz
(Factory
Default)
22.5 kHz
22.5 kHz
4.5 kHz
(Factory
Default)