![National Instruments SCC-CTR01 Скачать руководство пользователя страница 18](http://html1.mh-extra.com/html/national-instruments/scc-ctr01/scc-ctr01_user-manual-and-specifications_3701402018.webp)
NI SCC-CTR01 User Guide and Specifications
18
ni.com
Protection
Short circuit (on output pins)..................0.6 A min, 1.1 A max
(stays off after detecting a short
circuit and retries to operate every
250 ms, then automatically
recovers after removing the short)
Reverse and overvoltage
on output and V
dd
pins ............................±60 VDC max, continuous
Physical Characteristics
Dimensions .............................................8.89 cm
×
2.92 cm
×
1.85 cm
(3.50 in.
×
1.15 in.
×
0.73 in.)
Mass........................................................28.6 g (1.0 oz)
I/O connectors.........................................One 20-pin right-angle
male connector,
one 7-pin screw terminal
Field-wiring diameter .............................28 to 16 AWG
Environmental
Operating temperature ............................0 to 50 °C
Storage temperature ................................–20 to 70 °C
Humidity .................................................10 to 90% RH, noncondensing
Maximum altitude...................................2,000 m
Pollution Degree (indoor use only) ........2
Safety
This product is designed to meet the requirements of the following
standards of safety for electrical equipment for measurement, control, and
laboratory use:
•
IEC 61010-1, EN 61010-1
•
UL 61010-1, CSA 61010-1
Note
For UL and other safety certifications, refer to the product label or visit
ni.com/
certification
, search by model number or product line, and click the appropriate link
in the Certification column.