Table 5. DC Block Usage Guidelines
DC Block
Measurement Frequency
Maximum Allowed Voltage
Removed
≤16 kHz
0 V DC
Attached
≥16 kHz
-2 V DC to +2 V DC
Removing the DC Block
Figure 6. Removing the DC Block from the PXIe-5606 RF IN Connector
2
1
3
2
1. DC block
2. Screws
3. DC block support
1.
Remove the two screws on the DC block support with a Torx screwdriver with a T8 bit.
2.
Loosen the DC block support with a wrench and remove.
3.
Loosen and remove the DC block from the RF IN connector with a torque wrench.
PXIe-5668 Getting Started Guide
|
© National Instruments
|
15