38
|
ni.com
|
PXIe-5665 Calibration Procedure
13. Compare the
PXIe-5665 Average Noise (dBm/Hz)
to the verification test limits in Table 18,
Table 19, or Table 20 as appropriate.
Table 18.
Average Noise Verification Test Limits (Preamplifier Disabled, PXIe-5665
14 GHz VSA Preselector Disabled)
Frequency
As-Found Limit
(dBm/Hz)
As-Left Limit
*
(dBm/Hz)
>10 MHz to 100 MHz
-149
-151
>100 MHz to 300 MHz
-152
-154
>300 MHz to 1.7 GHz
-151
-153
>1.7 GHz to 2.8 GHz
-149
-151
>2.8 GHz to 3.6 GHz
-148
-150
>3.6 GHz to 7.5 GHz
†
-148
-150
>7.5 GHz to 8.5 GHz
†
-146
-147
>8.5 GHz to 12 GHz
†
-147
-148
>12 GHz to 14 GHz
†
-145
-146
*
section of this document for more information
about as-left limits.
†
Frequencies >3.6 GHz apply only to the PXIe-5665 14 GHz VSA.
Table 19.
Average Noise Verification Test Limits (Preamplifier Enabled)
Frequency
As-Found Limit
(dBm/Hz)
As-Left Limit
*
(dBm/Hz)
>10 MHz to 100 MHz
-161
-163
>100 MHz to 300 MHz
-162
-165
>300 MHz to 1.7 GHz
-162
-164
>1.7 GHz to 2.8 GHz
-161
-163
>2.8 GHz to 3.6 GHz
-160
-163
*
section of this document for more information
about as-left limits.